114, 115, 116, and 117
Calibration Information
2
Safety Information
"Warning" and "Caution" Statements
A “XW Warning" identifies hazardous conditions and actions that could cause bodily harm or death.
A "W Caution" identifies conditions and actions that could damage the Meter, the equipment under test, or cause
permanent loss of data.
XWWarnings and Precautions
To avoid possible electric shock or personal injury, follow these guidelines:
• Use the Meter only as specified in this manual or the protection provided by
the Meter might be impaired.
• Do not use the Meter or test leads if they appear damaged, or if the Meter is
not operating properly.
• Always use proper terminals, switch position, and range for measurements.
• Verify the Meter's operation by measuring a known voltage. If in doubt, have
the Meter serviced.
• Do not apply more than the rated voltage, as marked on Meter, between
terminals or between any terminal and earth ground.
• Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc.
These voltages pose a shock hazard.
• Disconnect circuit power and discharge all high-voltage capacitors before
testing resistance, continuity, diodes, or capacitance.
• Do not use the Meter around explosive gas or vapor.
• When using test leads or probes, keep your fingers behind the finger
guards.
• Remove test leads from Meter before opening the battery door or Meter
case.
• Comply with local and national safety requirements when working in
hazardous locations.
• Use proper protective equipment, as required by local or national authorities
when working in hazardous areas.
• Avoid working alone.
• Use only the replacement fuse specified or the protection may be impaired.
• Check the test leads for continuity before use. Do not use if the readings are
high or noisy.
• Do not use the Auto-V LoZ function to measure voltages in circuits that
could be damaged by this function’s low input impedance (≈3 kΩ) (114, 116
and 117).
Содержание
75- 115 116 and 117 calibration information
- Calibration information
- 115 116 and 117
- Contact information
- Introduction
- Digital multimeters
- Xw warnings and precautions
- Safety information
- Table 1 lists the international symbols that appear in this document and on the meter
- International electrical symbols
- General specifications
- Digital multimeters international electrical symbols
- Table 2 accuracy specifications
- 115 116 and 117 calibration information
- If the display shows any other value have the meter serviced see contact information earlier in this document
- If the display shows a resistance value in the range of that shown in figure 1 the fuse is good
- If the display reads 0l replace the fuse and test again
- Figure 1 fuse testing
- Digital multimeters basic maintenance
- Basic maintenance
- Testing the fuse 115 117 only to test the fuse
- Testing the fuse 115 117 only
- Table 3 input characteristics
- Set the rotary switch to
- Plug a test lead into the v jack and touch the probe to the 10a jack as shown in figure 1
- Xw warning
- Replacing the battery and fuse
- To avoid electric shock do not perform the performance test procedures unless the meter is fully assembled
- The following performance tests verify the complete operation of the meter and check the accuracy of each meter function against its specifications the recommended calibration interval is 12 months if the meter fails any part of the test calibration adjustment and or repair is indicated
- Required equipment table 4 lists the equipment required to conduct a performance test on the meter
- Required equipment
- Performance tests
- In the performance tests the meter is referred to as the unit under test uut
- Digital multimeters performance tests
- Cleaning the meter wipe the case with a damp cloth and mild detergent do not use abrasives or solvents dirt or moisture in the terminals can affect readings
- Cleaning the meter
- Xw warning
- Xw warning
- Testing the display
- Preparing for the performance tests
- Keypad test
- Backlight test
- Turn the rotary switch to the function listed in each step of table 5
- To verify the accuracy of the dmm functions do the following
- Testing temperature 116 only connect the k type thermocouple to the temperature input of the meter and temperature calibrator to ensure an accurate measurement the meter and the thermocouple connector must be at the same temperature after connecting the thermocouple to the meter allow the junctions to stabilize before recording the displayed reading this can take several minutes depending on temperature gradients
- Testing temperature 116 only
- If the display reading falls outside of the range shown in table 5 the meter requires calibration adjustment or repair
- Connect the calibrator to the v ω and com input terminals on the meter
- Compare the reading on the meter display with the display reading in table 5
- Check the fuses and battery and replace them if necessary refer to testing the fuses and replacing the battery and fuse
- Apply the input level for each step listed in table 5
- Allow the temperature of the uut to stabilize at room temperature 23 c 5 c 73 f 9 f
- Warm up the calibrator as required by its specifications
- Set calibrator to standby reconfigure leads and program for amps output
- 115 116 and 117 calibration information
- Testing the voltalert function 117 only
- Calibration adjustment password
- Calibration adjustment
- Changing the password
- Xw warning
- Restoring the default password
- Meter buttons used in the calibration steps
- Calibration adjustment procedure
- Notes set the calibrator to standby prior to changing the function switch position and after completing adjustment of each function
- If the calibration adjustment procedure is not properly completed the meter will not operate correctly
- Digital multimeters calibration adjustment
- 115 116 and 117 calibration information
- Digital multimeters replacement parts
- Table 7 lists the meter s replaceable parts identified in figure 6
- Replacement parts
- Figure 6 exploded view of meter
- 115 116 and 117 calibration information
- Warranty
- Digital multimeters warranty
- Battery primary mno2 zn 9v 505mah 6lr61 alkaline 17x26x48mm bulk
Похожие устройства
-
Fluke 15b+Инструкция РУС -
Fluke 28 IIИнструкция по эксплуатации -
Fluke 87V с поверкойИнструкция по эксплуатации -
Fluke 106 с поверкойИнструкция по эксплуатации -
Fluke 117Extended Specs -
Fluke 117Дополнение к информации о калибровке -
Fluke 117Положение об энергозависимости памяти -
Fluke 116СРуководство по эксплуатации -
Fluke 116Руководство по эксплуатации -
Fluke 116Информация о калибровке -
Fluke 116Add wireless to your Fluke digital multimeter -
Fluke 116DMM Selection Guide Brochure
Узнайте о важной информации по калибровке и мерам безопасности при использовании измерительных приборов. Соблюдайте рекомендации для предотвращения травм и повреждений.