Fluke 710 [15/20] Hart transmitter 709h 710 only
![Fluke 709 [15/20] Hart transmitter 709h 710 only](/views2/1435991/page15/bgf.png)
Precision Loop Calibrator
Performance Verification Tests
11
Hart Transmitter (709H/710 Only)
1. Push on the Product.
2. Turn the knob to select mA Measure with 24 V.
3. Connect the red jack to the + HART connection on a transmitter. Connect the black jack
to the – HART connection on a transmitter. See Figure 5.
hid015.eps
Figure 5. HART Transmitter Connections
4. Push on the Product.
5. Turn the knob to select Hart Comm. Menu and then push the rotary knob.
6. The display shows if the 250 Ω for HART is on or off. If the transmitter used does not
have a built in 250 Ω Hart resistor, the 250 Ω resistor will need to be on. Turn the rotary
knob to highlight the 250 Ω Resistor then push the rotary knob to change states if
necessary.
7. Turn the rotary knob to select Hart 250 Ω Connect and then push the rotary knob. If
properly connected, it will start polling for HART devices. When one is found, push the
rotary knob. If data acquisition succeeds, the HART resistor and loop supply are
operating correctly.
This concludes the performance verification tests.
Содержание
- 709h 710 1
- 709h 710 calibration manual 1
- Warranty 2
- Table of contents 3
- Contact fluke 5
- Introduction 5
- To prevent possible electrical shock fire or personal injury read all safety information before you use the product 5
- Xw warning 5
- Safety information 6
- Specifications 6
- Battery replacement 8
- Clean the product 8
- Maintenance 8
- Xw warning 8
- Precision loop calibrator 9
- 709h 710 10
- Performance verification tests 10
- Required test equipment 10
- Table 1 is a list of the required equipment for the performance verification tests and calibration adjustment 10
- The performance tests verify the complete operation of the product and measure the accuracy of each function against product specifications if the product fails a part of the test calibration adjustment or repair is necessary see calibration adjustment procedure 10
- The performance verification tests check the accuracy of each product function against its specifications if the product fails any of these tests calibration adjustment or repair is necessary 10
- To prevent possible electrical shock fire or personal injury do not perform the performance test or verification test procedures unless the product is fully assembled 10
- Xw warning 10
- Dc current source mode 11
- 709h 710 12
- A connect a test lead from the red aux terminal of the dc calibrator calibrator to the dmm i input 12
- B connect a test lead from the black aux terminal of the calibrator to the com terminal on the product 12
- C connect a test lead from the dmm lo input to the product terminal see figure 3 12
- Connect the product 12
- Dc current measurement mode 12
- Disconnect all connections to the product 12
- Make sure that the product shows source in the upper left corner 12
- Push the rotary knob 12
- See table 2 to verify the readings on the dmm for the subsequent tests no adjustment is necessary for test 1 12
- Set the dmm to measure dc current 12
- Test 2 push twice to select 12 00 ma verify the readings 12
- Test 3 push twice to select 20 00 ma push and rotate the knob until 24 00 ma is shown verify the readings 12
- The product shows measure 12
- To verify the performance of the dc current measurement function 12
- Turn the rotary knob to select ma measure 12
- Dc voltage measurement mode 14
- Hart transmitter 709h 710 only 15
- Calibration adjustment procedure 16
- Ma measure mode 16
- Ma source 17
- Voltage measure 18
- Precision loop calibrator 19
- User replaceable parts 19
- User replaceable parts are listed in table 3 and shown in figure 6 for more information about these items and their prices contact a fluke representative see the contact fluke section 19
- 709h 710 20
- Calibration manual 20
- Figure 6 user replaceable parts 20
Похожие устройства
- Fluke 710 Краткий справочник
- Fluke 710 Руководство по эксплуатации
- Fluke 789 Положение об энергозависимости памяти
- Fluke 789 Приложение к руководству по калибровке
- Fluke 789 Приложение к Руководству по эксплуатации
- Fluke 789 ProcessMeter
- Fluke 789 Паспорт безопасности
- Fluke 789 Краткий справочник
- Fluke 789 Руководство по калибровке
- Fluke 789 Руководство по эксплуатации
- Fluke 787B Положение об энергозависимости памяти
- Fluke 787B Приложение к руководству по калибровке
- Fluke 787B ProcessMeter
- Fluke 787B Приложение к Руководству по эксплуатации
- Fluke 787B Краткий справочник
- Fluke 787B Паспорт безопасности
- Fluke 787B Руководство по эксплуатации
- Fluke 787B Руководство по калибровке
- Fluke 787 Положение об энергозависимости памяти
- Fluke 787 Приложение к руководству по калибровке