Fluke 710 [6/20] Safety information
![Fluke 710 [6/20] Safety information](/views2/1435991/page6/bg6.png)
709/709H/710
Calibration Manual
2
Safety Information
For complete information about how to safely use this Product, read the 709/709H/710
Safety Information included with the Product or located on the Fluke website.
Specifications
Ranges
mA ........................................................ 0 mA to 24 mA
Volts ..................................................... 0 V dc to 30 V dc
Resolution
mA Ranges .......................................... 1 μA
Voltage Range ..................................... 1 mV
Accuracy ..................................................... 0.01 % ±2 LSD all ranges (@23 °C ±5 °C)
Stability 20 ppm of F.S. /°C from -10 °C to 18 °C and
28 °C to 55 °C
Operating Temp Range .............................. -10 °C to 50 °C (14 °F to 122 °F)
Storage Temp Range ................................. -20 °C to 60 °C (-4.0 °F to 140 °F)
Altitude ........................................................ 3000 meters
Ingress Protection Rating ........................... IEC 60529: IP40
Humidity Range .......................................... 10 % to 95 % non-condensing
Display 128 x 64 pixels, LCD Graphic with backlight,
8.6 mm high digits
Power .......................................................... Six IEC LR03 batteries
Battery Life (Alkaline Batteries) .................. ≥40 hours continuous use (measure mode)
Loop Compliance Voltage .......................... 24 V dc @ 20 mA
Loop Drive Capability 1200 Ω without HART resistor, 950 Ω with
HART resistor
Electromagnetic Environment..................... IEC 61326-1 (portable equipment)
Dimensions (LxWxD) .................................. 152 x 93 x 44 mm, (6.0 x 3.7 x 1.7 in)
Weight ......................................................... 0.3 kg (9.5 oz)
Содержание
- 709h 710 1
- 709h 710 calibration manual 1
- Warranty 2
- Table of contents 3
- Contact fluke 5
- Introduction 5
- To prevent possible electrical shock fire or personal injury read all safety information before you use the product 5
- Xw warning 5
- Safety information 6
- Specifications 6
- Battery replacement 8
- Clean the product 8
- Maintenance 8
- Xw warning 8
- Precision loop calibrator 9
- 709h 710 10
- Performance verification tests 10
- Required test equipment 10
- Table 1 is a list of the required equipment for the performance verification tests and calibration adjustment 10
- The performance tests verify the complete operation of the product and measure the accuracy of each function against product specifications if the product fails a part of the test calibration adjustment or repair is necessary see calibration adjustment procedure 10
- The performance verification tests check the accuracy of each product function against its specifications if the product fails any of these tests calibration adjustment or repair is necessary 10
- To prevent possible electrical shock fire or personal injury do not perform the performance test or verification test procedures unless the product is fully assembled 10
- Xw warning 10
- Dc current source mode 11
- 709h 710 12
- A connect a test lead from the red aux terminal of the dc calibrator calibrator to the dmm i input 12
- B connect a test lead from the black aux terminal of the calibrator to the com terminal on the product 12
- C connect a test lead from the dmm lo input to the product terminal see figure 3 12
- Connect the product 12
- Dc current measurement mode 12
- Disconnect all connections to the product 12
- Make sure that the product shows source in the upper left corner 12
- Push the rotary knob 12
- See table 2 to verify the readings on the dmm for the subsequent tests no adjustment is necessary for test 1 12
- Set the dmm to measure dc current 12
- Test 2 push twice to select 12 00 ma verify the readings 12
- Test 3 push twice to select 20 00 ma push and rotate the knob until 24 00 ma is shown verify the readings 12
- The product shows measure 12
- To verify the performance of the dc current measurement function 12
- Turn the rotary knob to select ma measure 12
- Dc voltage measurement mode 14
- Hart transmitter 709h 710 only 15
- Calibration adjustment procedure 16
- Ma measure mode 16
- Ma source 17
- Voltage measure 18
- Precision loop calibrator 19
- User replaceable parts 19
- User replaceable parts are listed in table 3 and shown in figure 6 for more information about these items and their prices contact a fluke representative see the contact fluke section 19
- 709h 710 20
- Calibration manual 20
- Figure 6 user replaceable parts 20
Похожие устройства
- Fluke 710 Краткий справочник
- Fluke 710 Руководство по эксплуатации
- Fluke 789 Положение об энергозависимости памяти
- Fluke 789 Приложение к руководству по калибровке
- Fluke 789 Приложение к Руководству по эксплуатации
- Fluke 789 ProcessMeter
- Fluke 789 Паспорт безопасности
- Fluke 789 Краткий справочник
- Fluke 789 Руководство по калибровке
- Fluke 789 Руководство по эксплуатации
- Fluke 787B Положение об энергозависимости памяти
- Fluke 787B Приложение к руководству по калибровке
- Fluke 787B ProcessMeter
- Fluke 787B Приложение к Руководству по эксплуатации
- Fluke 787B Краткий справочник
- Fluke 787B Паспорт безопасности
- Fluke 787B Руководство по эксплуатации
- Fluke 787B Руководство по калибровке
- Fluke 787 Положение об энергозависимости памяти
- Fluke 787 Приложение к руководству по калибровке