Fluke 787B [12/44] General specifications

Fluke 789 [12/44] General specifications
789/787B
Calibration Manual
8
General Specifications
Maximum Voltage between
any Terminal and Earth Ground .......................... 1000 V
Fuse Protection for
mA inputs ............................................................... 0.44 A, 1000 V IR 10 kA
Power
Battery Type ........................................................ IEC LR6 (AA Alkaline)
Quantity ............................................................... 4
Temperature
Operating ............................................................ -20 °C to +55 °C
Storage ............................................................... -40 °C to +60 °C
Altitude
Operating ............................................................ 2000 m
Storage ............................................................... 12 000 m
Frequency Overload Protection ........................... 10
6
V Hz max
Temperature coefficient
Measurements .................................................... 0.05 x specified accuracy per °C for temperatures <18 °C or >28 °C
Source................................................................. 0.1 x specified accuracy per °C for temperatures <18 °C or >28 °C
Relative humidity .................................................. 95 % up to 30 °C, 75 % up to 40 °C, 45 % up to 50 °C, and 35 % up to
55 °C
Vibration ................................................................ Random 2g, 5 to 500 Hz
Shock ..................................................................... 1 meter drop test
Size ......................................................................... 10.0 cm X 20.3 cm X 5.0 cm (3.94 in X 8.00 in X 1.97 in)
Weight .................................................................... 610 g (1.6 lb)
Safety
General ............................................................... IEC 61010-1: Pollution Degree 2
Measurement ...................................................... IEC 61010-2-033: CAT IV 600 V / CAT III 1000 V
Electromagnetic Compatibility (EMC) ................. Accuracy for all ProcessMeter functions is not specified in RF field
>3 V/m
International ........................................................ IEC 61326-1: Portable Electromagnetic Environment; IEC 61326-2-2
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for
the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low-voltage power
supply network that supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated
disturbances.
Caution: This equipment is not intended for use in residential
environments and may not provide adequate protection to radio
reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.
Korea (KCC) ....................................................... Class A Equipment (Industrial Broadcasting & Communication
Equipment)
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user should take
notice of it. This equipment is intended for use in business
environments and not to be used in homes.
USA (FCC) .......................................................... 47 CFR 15 subpart B. This product is considered an exempt device per
clause 15.103.