Fluke 717 [5/5] Change 4
![Fluke 717 [5/5] Change 4](/views2/1436043/page5/bg5.png)
Supplement 717 Instruction.Sheet
4 8/17
Change #4, 67386
On the second panel, under Symbols, add:
Conforms to relevant South Korean
EMC Standards.
Under General Specifications, add:
Electromagnetic Compatibility (EMC)
International .......... IEC 61326-1: Portable Electromagnetic
Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary
for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments
other than domestic and those directly connected to a low-
voltage power supply network that supplies buildings used for
domestic purposes. There may be potential difficulties in
ensuring electromagnetic compatibility in other environments
due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential
environments and may not provide adequate protection to radio
reception in such environments.
Emissions that exceed the levels required by CISPR 11 can
occur when the equipment is connected to a test object.
Korea (KCC) ......... Class A Equipment (Industrial Broadcasting
& Communication Equipment)
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user should
take notice of it. This equipment is intended for use in business
environments and not to be used in homes.
USA (FCC) ......... 47 CFR 15 subpart B. This product is
considered an exempt device per clause
15.103.
Содержание
- Supplement 1
- Change 1 2
- Change 2 2
- Supplement 717 instruction sheet 3
- Under specifications replace the entire section with specifications 3
- And power on the calibrator 4
- Change 3 4
- In the specification section add the following 4
- Instruction sheet supplement 4
- Specifications apply to the 717 pressure calibrators with firmware version 4 or greater to verify the firmware version push and hold 4
- Change 4 5
- Conforms to relevant south korean emc standards 5
- On the second panel under symbols add 5
- Supplement 717 instruction sheet 5
- Under general specifications add 5
Похожие устройства
- Fluke 717 Руководство по калибровке
- Fluke 714 Приложение к руководству по калибровке
- Fluke 714 Руководство по калибровке
- Fluke 712 Приложение к руководству по калибровке
- Fluke 712 Руководство по калибровке
- Fluke 718 Приложение к руководству по калибровке
- Fluke 718 Руководство по калибровке
- Fluke 715 Приложение к руководству по калибровке
- Fluke 715 Руководство по калибровке
- Fluke 707 ex Положение об энергозависимости памяти
- Fluke 707 ex 707Ex CCD _ Контрольные схемы
- Fluke 707 ex Дополнение к информации о калибровке
- Fluke 707 ex Информация о калибровке
- Fluke 707 ex Приложение к паспорту безопасности
- Fluke 707 ex Паспорт безопасности
- Fluke 707 ex Приложение к Руководству по эксплуатации
- Fluke 705 Дополнение к информации о калибровке
- Fluke 705 Приложение к Инструкции
- Fluke 705 Приложение к паспорту безопасности
- Fluke 705 Паспорт безопасности