Fluke 707 ex [6/24] Battery saver
![Fluke 707 ex [6/24] Battery saver](/views2/1005454/page6/bg6.png)
707Ex
Users Manual
2
Battery Saver
The Calibrator automatically turns off after 30 minutes of
inactivity. To reduce this time or disable this feature:
1. With the Calibrator OFF, press D.
P.S. xx is displayed, where xx is the turn-off time in
minutes. OFF means the power saver is disabled.
2. Turn k to decrease or lto increase the turn-off time.
To disable, turn k until the display shows OFF.
3. The Calibrator resumes normal operation after
2 seconds.
Safety Information
In this manual a Warning identifies conditions and actions
that pose hazards to the user. A Caution identifies
conditions and actions that may damage the Calibrator or
equipment under test. International symbols used in this
manual are identified later in the Symbols section.
Read the entire Users Manual and the 707Ex mA
Calibrator CCD (Concept Control Drawing) before using
the Calibrator.
WWarnings and Cautions
To avoid electric shock, injury, or damage to the
Calibrator:
• Use the Calibrator only as described in this
Users Manual and the Fluke 707Ex mA
Calibrator CCD (Concept Control Drawing) or
the protection provided by the Calibrator may
be impaired.
• Inspect the Calibrator before use. Do not use it
if it appears damaged.
• Check the test leads for continuity, damaged
insulation, or exposed metal. Replace damaged
test leads.
• Never apply more than 28 V between the input
terminals, or between any terminal and earth
ground.
Applying more than 28 volts to the input
terminals invalidates the Calibrator’s Ex
Approval and may result in permanent damage
to the unit so it can no longer be used.
• Use the proper terminals, mode, and range for
your measuring or sourcing application.
• To prevent damage to the unit under test, put
the Calibrator in the correct mode before
connecting the test leads.
• When making connections, connect the COM
test probe before the live test probe. When
disconnecting, disconnect the live probe before
the COM probe.
Содержание
- Ex ma calibrator 1
- Ma calibrator 1
- March 2003 rev 3 5 03 english 1
- Users manual 1
- Table of contents 3
- Title page 3
- Introduction 5
- Ma calibrator 5
- Warning 5
- Battery saver 6
- Safety information 6
- Warnings and cautions 6
- Faults and damage 7
- Safety advice 7
- Ex certification data 8
- Safety regulations 8
- Symbols 8
- Front panel controls 9
- Changing the ma output span 10
- Sourcing ma 10
- Using the ma sourcing output modes 10
- Auto ramping the ma output 11
- Simulating a transmitter 11
- Measuring dc ma 12
- Using the spancheck function 12
- Measuring dc ma with loop power 13
- Hart compatibility 14
- Measuring dc volts 14
- Calibration 15
- Cleaning 15
- In case of difficulty 15
- Maintenance 15
- Replacing the battery 16
- Approved batteries 17
- Fluke part numbers 17
- Replaceable parts 17
- Accuracy specifications 18
- General specifications 18
- Contacting fluke 20
- Zelm ex 21
- Fm certificate 23
Похожие устройства
- Supra SNP-431 Инструкция по эксплуатации
- Digitech BNX3 Инструкция по эксплуатации
- LG MH-6646QMB Инструкция по эксплуатации
- Hotpoint-Ariston ARUSL 105 Инструкция по эксплуатации
- Geo-Fennel FL 200 A-N 291000 Инструкция по эксплуатации
- Fluke 712 Инструкция по эксплуатации
- Digitech BP 200 Инструкция по эксплуатации
- Supra SNB-200 Инструкция по эксплуатации
- LG MH-6646QM Инструкция по эксплуатации
- Hotpoint-Ariston ARSL 105 (CSI).L Инструкция по эксплуатации
- Geo-Fennel FL 55 Multilinner 500050 Инструкция по эксплуатации
- Fluke 714 Инструкция по эксплуатации
- Digitech GNX2 Инструкция по эксплуатации
- Supra SNP-351 Инструкция по эксплуатации
- Belkin F8N136eaKSG Инструкция по эксплуатации
- LG MH-6643C Инструкция по эксплуатации
- Hotpoint-Ariston ARSF 125 CSI.L Инструкция по эксплуатации
- Geo-Fennel FLG 210 A green 291700 Инструкция по эксплуатации
- Digitech GNX3 Инструкция по эксплуатации
- Centurion COMFORT Инструкция по эксплуатации