IKA ICC basic eco 18 c 0008035800 [4/37] Intelligence

IKA ICC control eco 8 c 0008036100 [4/37] Intelligence
Safety
/// All IKA tempering products meet the highest safety standards
All of the IKA temperature control devices that are equipped with a
heating function meet the highest safety classification III (FL) for use
with combustible fluids pursuant to DIN 12876.
Safety temperature
The safety temperature can be adjusted
using tools or through the display. The
temperature is monitored by an indepen-
dent temperature sensor.
Thermofluid temperature threshold
The thermal fluid that is being used can
be selected in the menu. This ensures
that the temperature remains outside the
critical values for that particular fluid.
Minimum and maximum temperature
values can be adjusted manually within
these parameters.
Speed
The speed can be restricted; which allows
for determining the maximum pump
pressure.
Carrying handle
Safe carrying and positioning (ICC).
Recessed handles
Ergonomic handling (HBC and RC 2).
Bracket
Secures the base and protects the
floats and tubular heater (ICC).
Casters
Facilitate easy positioning of the device
(RC, HBC, CBC).
Transport handle
Easy and safe handling (HBC).
Safe and complete drainage of baths
The bath can be fully emptied of thermal
fluids, in a simple and clean manner. The
physical separation of the drain valve and
the opening screw ensures that the user
does not come into contact with the fluid.
SAFE HANDLING DUE TO ERGONOMIC AND WELL THOUGHT-OUT DESIGNS
VISUAL AND AUDIBLE ALARM
Indication of critical fluid level, critical
temperature or blocked pump.
LOCK FUNCTION
Locks set parameters to prevent accidental
adjustments on the WiCo.
FLUID LEVEL DETECTION
Critical minimum or maximum levels are re-
cognized both mechanically (by the floater)
as well as electronically (by a temperature
sensor).
SAFE OPERATION DUE TO
ADJUSTABLE THRESHOLD VALUES
Power
/// Superior power
Whether infinitely adjustable pressure/suction pump or premium
grade stainless steel: Rely on a strong partner.
Tempering
The IKA heating and refrigerating circu-
lators temper fluids ranging from - 30 °C
to + 250 °C with a heating and cooling
power of up to 2.5 kW and 1.4 kW
respectively.
For decades, temperature
control has been one of IKA's
core competencies
IKA heating circulators reach a tem-
perature constancy of up to ± 0.01 K.
The power-regulated compressor of the
recirculating chiller allows a constant
temperature of up to 0.05 K.
The large heating surfaces gently control
the temperature of the thermal fluids
and ensure outstanding heat transfer.
The strong heat output of the circulators
ensures short heat-up times.
All IKA heating circulators have the option
of adding a cooling coil with large cooling
surface for operating in or below ambient
temperature. By connecting a chiller, it is
possible to reach an operating tempera-
ture down to - 20 °C.
Pressure/suction pump
All IKA circulators use a powerful, infinite-
ly adjustable pressure/suction pump made
of high-quality PEEK plastic. It allows
for flexible use of the devices in open or
closed system applications. The pump en-
sures homogenous mixing inside the bath
and provides a high flow rate for external
applications.
All circulators are equipped or can be
retrofitted using pump connections
(M 16 × 1).
Pump curve
The actual maximum flow rate can be
determined under known pressure loss of
the experiment based on a pump curve.
Pressure losses occur, for instance, from
height differences, narrow and long hoses
as well as the high viscosity of a bath
fluid. The maximum viscosity of the bath
fluid is 50 mPas.
Energy eciency
The excellent insulation and the demand-
driven output control ensure that IKA
temperature control instruments are
highly energy-ecient.
As such, IKA recirculating chillers use up
to 60 percent less energy than compa-
rable competitors' devices in standard
applications.
Robust and durable
IKA temperature control products are
made of high-quality materials and
designed for a long service life.
Parts that come into contact with prod-
ucts are made of premium grade stainless
steel (V4A) and highly durable PEEK, FKM
and PTFE; meeting the basic requirements
for use in the food industry.
Pressure/suction pump
Intelligence
/// Intuitive operation and automatic tempering
Smart technology as well as user-friendly menu navigation simplify
temperature control for any application.
Operating modes
There are four options after a restart or
power failure – simply select the mode
of operation.
Operating mode A:
After switching on/power interruption,
no automatic restart of functions.
Operating mode B:
After switching on/power interruption,
automatic start of functions depending
on the previous settings. Ideal when
working via switchable sockets.
Operating mode C:
Set values (set in A or B) cannot be
modified. After switching on/power
interruption, automatic start of functions
depending on the previous settings.
Operating mode D:
Confirmation query for set value changes;
if functions are active. After switching on/
power interruption, no automatic restart
of functions.
Intuitive operation
User-friendly menu navigation,
push buttons and knobs enable
simple operation.
Connectivity
USB and RS 232 are standard interfaces.
Software programs are used to gather the
measured data and control the devices,
e.g. labworldsoft
®
by IKA. Following
registration, the Firmware Update Tool
ensures that users always have the latest
version of the software.
Most IKA circulators have a PT 100
interface for the display and control
of the external temperature (except for
the ICC basic, RC 2 | 5 basic and HRC 2
basic ).
Calibration and adjustment
The internal (and, if available external)
temperature sensor can be adjusted
either via a two- or three-point calibration
process.
Automatic tempering
Before starting the tempering process,
the control parameters of the bath fluid
as well as the amount of the bath fluid
are measured in order to prevent the
temperature from being exceeded. There
are also options to manually set the PID
control parameters.
Software control/specification
of heating rates
Labworldsoft
®
software can be used to
precisely specify temperature ramps and
heat-up times/heating rates.
7
9
Temperature control productsTemperature control products

Содержание

Скачать