Fluke 724 [41/56] Auto ramping the output
![Fluke 724 [41/56] Auto ramping the output](/views2/1005521/page41/bg29.png)
Temperature Calibrator
Storing and Recalling Setups
31
Auto Ramping the Output
Auto ramping gives you the ability to continuously apply a
varying stimulus from the calibrator to a transmitter, while
your hands remain free to test the response of the
transmitter.
When you press
L, the calibrator produces a
continuously repeating 0 % - 100 % - 0 % ramp in your
choice of three ramp waveforms:
•
E 0 % - 100 % - 0 % 40-second smooth ramp
• P 0 % - 100 % - 0 % 15-second smooth ramp
• N 0 % - 100 % - 0 % Stair-step ramp in 25 %
steps, pausing 5 seconds at each step.
To exit ramping, press any button.
Storing and Recalling Setups
You can store up to eight of your settings in a nonvolatile
memory and recall the settings for later use. A low battery
condition or a battery change does not jeopardize the
stored settings. Proceed as follows:
1. After you create a calibrator setup, press
S. In the
display, the memory locations appear.
2. Press
Y or Z to select locations one through eight. An
underscore appears below the selected memory
location.
3. Press
S. Only the stored memory location will be
displayed. The setup is stored.
To recall setups, proceed as follows.
1. Press
Q. The memory locations appear on the
display.
2. Press
Y or Z to select the appropriate location and
press
Q.
Содержание
- February 2000 rev 8 03 1
- Temperature calibrator 1
- Users manual 1
- Limited warranty limitation of liability 2
- Table of contents 3
- Title page 3
- Temperature calibrator contents 5
- List of tables 7
- Table title page 7
- Figure title page 9
- List of figures 9
- Contacting fluke 11
- Introduction 11
- Temperature calibrator 11
- Safety information 13
- Standard equipment 13
- Warning 14
- Caution 15
- Warning 15
- Alligator clips 16
- Figure 1 standard equipment 16
- Stackable test leads 16
- Tl75 test lead set 16
- Users manual 16
- Zi01f eps 16
- 6 5 4 3 2 18
- Figure 2 input output terminals and connectors 18
- Figure 2 shows the calibrator input and output terminals table 3 explains their use 18
- Getting acquainted with the calibrator 18
- Input and output terminals 18
- Source 18
- Tc rtd 18
- Users manual 18
- Figure 3 keys 20
- Figure 3 shows the calibrator keys and table 4 explains their use 20
- Users manual 20
- Display 23
- Getting started 24
- Shut down mode 24
- Figure 5 voltage to voltage test 25
- Recall 25
- Source 25
- Tc rtd 25
- Temperature calibrator shut down mode 25
- Zi04f eps 25
- Contrast adjustment 26
- Current measurement with loop power 27
- Measuring electrical parameters upper display 27
- Using measure mode 27
- Figure 8 connections for supplying loop power 28
- Source 28
- Tc rtd 28
- Users manual 28
- Measuring electrical parameters lower display 29
- Measuring temperature 30
- Using thermocouples 30
- Figure 10 measuring temperature with a thermocouple 32
- Process temperature 32
- Tc miniplug 32
- Users manual 32
- Warning 30 v maximum to 32
- Using resistance temperature detectors rtds 33
- Figure 11 measuring temperature with an rtd measuring 2 3 and 4 wire resistance 35
- Resistance to be measured 35
- Temperature calibrator using measure mode 35
- Sourcing electrical parameters 36
- Using source mode 36
- Simulating rtds 37
- Simulating thermocouples 37
- Color depends on type of tc 38
- Figure 13 connections for simulating a thermocouple 38
- Source 38
- Tc miniplug 38
- Tc rtd 38
- Users manual 38
- Figure 14 connections for simulating 3 wire rtd 39
- Recall 39
- Sensor terminals 39
- Source 39
- Tc rtd 39
- Temperature calibrator using source mode 39
- Manually stepping the output 40
- Setting 0 and 100 output parameters 40
- Stepping and ramping the output 40
- Auto ramping the output 41
- Storing and recalling setups 41
- Calibrating a transmitter 42
- Figure 15 calibrating a thermocouple transmitter 43
- Source 43
- Tc rtd 43
- Temperature calibrator calibrating a transmitter 43
- Testing an output device 44
- Is good problems while measuring using the right jacks indicate that f3 may have opened to replace the fuse refer to figure 17 and perform the following steps 45
- Replacing the batteries 45
- Replacing the fuse 45
- The fuse can be removed and checked for resistance a value of 10 45
- Turn the calibrator off remove the test leads from the terminals and hold the calibrator face dow 45
- Warning 45
- Cleaning the calibrator 46
- Maintenance 46
- Service center calibration or repair 46
- Replacement parts 47
- Figure 18 replacement parts 48
- Users manual 48
- Dc ma measurement 49
- Dc voltage measurement 49
- Dc voltage source 49
- Specifications 49
- Ohms measurement 50
- Ohms source 50
- Millivolt measurement and source 51
- Temperature thermocouples 51
- Temperature rtd ranges and accuracies its 90 52
- General specifications 53
- Loop power supply 53
Похожие устройства
- Condtrol UBEXI XC 500 D Инструкция по эксплуатации
- Centurion XP v.1 Инструкция по эксплуатации
- LG MH-6354G Инструкция по эксплуатации
- Belkin F8N372cwBKW Инструкция по эксплуатации
- Digitech RP80 Инструкция по эксплуатации
- Candy GOY 1054 L-07 Инструкция по эксплуатации
- Fluke 725 Инструкция по эксплуатации
- Telwin Superior Plasma 60 HF 400V Инструкция по эксплуатации
- Centurion BIKE KEEPER Инструкция по эксплуатации
- Digitech S100 Инструкция по эксплуатации
- LG MH-6352U Инструкция по эксплуатации
- Candy GO4 107 DF 07 Инструкция по эксплуатации
- Fluke 725 ex Инструкция по эксплуатации
- Telwin SUPERPlasma 80/3 HF Инструкция по эксплуатации
- LG MH-6352T Инструкция по эксплуатации
- Digitech S200 Инструкция по эксплуатации
- Belkin F8N337cw Black Инструкция по эксплуатации
- Centurion BIKE Инструкция по эксплуатации
- Candy GO4 1064 D 07S Инструкция по эксплуатации
- Fluke 726 Инструкция по эксплуатации