Beurer GL40 (mg/dL) [33/40] Troubleshooting

Beurer GL40 (mg/dL) [33/40] Troubleshooting
Beurer GL40 mg/dL 33
8 Troubleshooting
Messages on the display for batteries and blood sugar measurement
No. Cause Action
LP Flat batteries Replace all batteries.
Ht
Temperature of the measurement
room, measuring device or test strip
above the allowable limit.
Repeat test with new test strip as soon as
the measurement room, measuring device
and test strip have reached room tempera-
ture (+20°C to +25°C).
Lt
Temperature of the measurement
room, measuring device or test strip
below the allowable limit.
Repeat test with new test strip as soon as
the measurement room, measuring device
and test strip have reached room tempera-
ture (+20°C to +25°C).
Err Used or contaminated
test strip was inserted.
Insert test strip which has not been
used or spoiled.
Repeat blood sugar measurement.
001 System error
Remove the batteries and reinsert
them. Contact our customer service
department if the problem persists.
005 Memory error
Remove the batteries and reinsert
them. Contact our customer service
department if the problem persists.