36 Beurer GL40 mg/dL
EMC
This unit complies with European standard EN 61326 and is subject to specific safety
precautions in terms of electromagnetic compatibility. Note that portable and mobile RF
communication equipment can affect this unit. Further details can be requested by con-
tacting the customer service address provided below.
Note regarding function of test strips
The test strips enable a quantitative measurement of the glucose content of capillary
whole blood. When the slit for collecting blood comes into contact with a drop of blood, it
fills automatically through simple capillary action. The blood is sucked into the absorbing
slit of the test strip and the measuring device measures the sugar level of the blood.
The test is based on the measurement of an electric current caused by a chemical reac-
tion between the glucose and the test strip reagent. The measuring device analyses this
current. The flow of current depends on the glucose content of the blood sample. The
results are shown in the display of the measuring device. Only a small quantity of blood is
required (0.6 microlitres) and the measurement takes approx. five seconds. The test
strips can detect blood sugar levels in the range from 20 to 630 mg/dL.
Chemical components of the test strip sensor
Glucose oxidase 0.6 IU
Electron shuttle 0.03 mg
Non-reactive components 0.06 mg