Beurer GL40 (mmol/L) Инструкция по эксплуатации онлайн [23/40] 54834

Beurer GL40 (mmol/L) Инструкция по эксплуатации онлайн [23/40] 54834
Beurer GL40 mmol/L 23
5.3 Evaluating a blood sugar measured value
Your blood sugar measuring device can process measured values between 1,1 and 35
mmol/L. The warning message "Lo" is displayed for measured values below 1,1 mmol/L.
The warning message "Hi" is displayed for a measured value above 35 mmol/L.
Note
If you suspect that the blood sugar values are incorrect, first repeat the test and, if
necessary, perform a function test using the control solution. If you consistently have
reason to doubt the results, contact your doctor.
Blood sugar
The following table shows a classification of blood sugar values according to the guide-
lines of the German Diabetes Association (DDG).
Time of blood sugar meas-
urement
Normal blood
sugar values
Suspicious Diabetes
On an empty stomach below 5,6 mmol/L 5,6-6,1 mmol/L Above 6,1 mmol/L
2 hours after eating below 7,8 mmol/L 7,8-11,1 mmol/L Above 11,1 mmol/L
Source: Deutsche Diabetes Gesellschaft (DDG) 2008
Скачать