ZCT 777
MEASUREMENT FOUNDATION
ZCT 777
16
“Two-point calibration”
Make the rst-point calibration. But after the adjustment the required foil thickness, make two-point calibration*.
The selection of the foil: the results of the following measurements must be between values of the foil 1 and 2. The value of
the foil 2 must be in approximately 1,5 times more than the value of the foil 1.
Shot-blased surfaces
The physical nature of shot-blased surfaces results in coating thickness reading that are too high. The mean thickness over
the peaks can be determined as follows (not that the statistics program is of great benet in this procedure):
Method A:
1.The gauge should be calibrated according One-point or Two point calibration. Use a smooth calibration sample with the
same curvature radius and the same substrate as the later measuring sample.
2.Now take approx. 10 readings on the uncoated, shot-blased sample to produce the mean value Xo
3.After this take approx. 10 further readings on coated, shot blasted test sample to produce the mean value Xm
4.The difference between the two mean values is the mean coating thickness Xeff over the peaks. The greater standard
deviation s of the two values Xm and Xo should also be taken into consideration:
Xeff=(Xm-Xo)±S, where S - is the measurement uncertainty of the value Xm and Xo.
Method B:
1.Carry out a zero calibration of 10 readings on a shot-blased, uncoated sample. Then carry out a foil calibration on the
uncoated substrate. The foil set should consist of a number of individual foils of max. 50 microns thickness each and should
roughly correspond to the estimated coating thickness.
2.The coating thickness can be read directly from display and should be averaged from 5….10 single measurements.