Ada ZCТ 777 [17/46] Zct 777

Ada ZCТ 777 [17/46] Zct 777
ZCT 777
ZCT 777
MEASUREMENT FOUNDATION
ZCT 777
17
Method C:
Simply follow the two-point calibration method using two foils as described in section Two-point calibration. The mean coat-
ing thickness should be calculated from 5…10 readings.
Measurements
After careful calibration has been made, all subsequent measurements will lie within the guaranteed measuring tolerance.
Example:
Readings: 150um, 156um, 153um
Mean value: X=153um
Standard deviation: s=±3um
Measuring uncertainty: u=±(1% of reading + 1um)
D= 153±3±(1,53um+1um) =153±5,5um
Gauge Control via PC
All measuring readings of all work modes can be downloaded to PC via USB port for data analysis. Setup software on your
PC: trigger le setup.exe
Trouble shooting (disk comes with the instrument). Trigger application THICKNESS from the shortcut on your desktop or
use menu “Launch”. You will see application window. Menu items (except for «Connect» and «Exit») are inactive until you
connect the instrument to the PC. After the connection with the cable (comes with the instrument) all menu items are active
(except for «Connect») (pic.1).