CHAUVIN ARNOUX CA-5005-MN89 [15/44] Resistances

CHAUVIN ARNOUX CA-5005-MN89 [15/44] Resistances
15
A~ with
3 A 10 A 30 A 100 A 300 A
(1)
miniclamp MN 89
Scale
(2)
3 10 3 10 3
Reading cœfficient x 1 x 1 x 10 x 10 x 100
Accuracy
(3)
10% 5% 2.5%
(4)
Protection ensured by the miniclamp MN89
(5)
(1) Range limitedto240AAC max by the miniclamp MN 89
(2) Marked with "
"
(3) In %ofthaend of scale
(4) Up to 200 A AC nominal range of the MN 89 miniclamp
(5) Measuring 240 A AC is equivalent to measuring 24 V DC
To get the resistance in , multiply the reading on the scale
(green) by the cœfficient of the selected range: x 1 or x 100.
x 1
(1)
x 100
Measurement 5 to 500 to
extend 10 k 1 M
Internal resistance 140 14 k
End of scale current 10 mA 100 µA
Accuracy
(2)
10%
Permitted overload 440 V
(1) Common to the continuily sound test function
(2) In % at mid-scale
7 - CONTINUITY SOUND TEST
Connection and specifications: idem Resistances
Place the selector switch on the
x 1 function
Continuous audible beep emitted for a resistance R < 50
6 - RESISTANCES
Never test a resistance on a live circuit.
If the “Voltest” light comes on, presence of a voltage U > 15 V
AC/DC. Unplug the leads immediately and cut off the voltage
before making the resistance measurement.
Connect to the terminals of the circuit or the component to test.
8 - GENERAL SPECIFICATIONS
8-1 Dimensions and weight
56 x 105 x 160 mm 500 g