B-4 Alternating Current Current (AC Current)
200µA
2mA
20mA
200mA
2A
10A
20A
0.1µA
1µA
10µA
100 µA
1mA
Range
Resolution
UT51 UT 52 UT53 UT54 UT55
Accuracy
(1%+3)
(1%+3)
10mA
(1.8%+3)
(3%+7)
(3%+7 )
(1.8%+3)
(1.8%+3)
Overload protection:
For UT51:
2A, 250V fast acting fuse, φ5x20mm(below 2A
range)
10A, 250V fast acting fuse,φ5x20mm(at 10A range) .
For UT52/53/54/55:
315mA, 250V fast acting fuse,φ5x20mm (No fuse
at 20A range).
Max current input :
For UT51: 10A (The measurement time for high
current should be less than 10 seconds maximum,
and the interval time between two measurements
should be greater than 15 minutes.
For UT52/53/54/55: 20A (The measurement time
for high current should be less than 15 seconds
maximum, and the interval time between two
measurements should be greater than 15 minutes.
Measuring voltage drop: Full range is 200mV.
Display: Average Value (RMS of Sine Wave ) .
Model UT51-55: OPERATING MANUAL