CHAUVIN ARNOUX CA-MN77 [12/32] Maintenance

CHAUVIN ARNOUX CA-MN93 [12/32] Maintenance
12
12
4.4.2 Frequency
n Use : 40 Hz to 10 kHz
Limit to 1 kHz if permanent operation at 200 A~.
4.4.3 Environmental conditions
n Indoor use
n Altitude: < 2000 m
n Climatic conditions: -10 to +55° C and RH < 85%
n Avoid splashing with water
4.5 Dimensions and weight
n Overall dimensions : 135 x 50 x 30 mm
n Weight: approx. 180 g
n Jaw opening : 21 mm
n Open jaw height : 69 mm
n Max. clamping capacity : Ø 20 mm cable or 20 x 5 mm bar.
4.6 Compliance with international norms
4.6.1 Electrical safety (as per IEC 61010-1 and 61010-2-032)
n Dual insulation n Pollution level 2
n Installation category III n Operating voltage 600
4.6.2 EC-compliant electromagnetic compatibility
n Emissions and immunity in an industrial setting compliant with EN61326-1.
4.6.3 Mechanical protection
n IP40 protection rating (as per EN60529) with jaws closed and IP30 with jaws
open.
4.6.4 Auto-extinction
n Jaws : V0 (as per UL 94)
5. MAINTENANCE
Only use specified spare parts for maintenance purposes. The
manufacturer cannot accept any responsibility for accidents occurring
following repairs carried out outside its after-sales department or approved
maintenance network.
5.1 Cleaning
The clamp must not be clamped to a cable and must be disconnected from the
measuring device. Do not splash water onto the clamp.
n Keep the jaw gap perfectly clean. Remove dust with a dry, soft cloth. Wipe the
iron jaws from time to time with an oil soaked cloth, in order to prevent rust
from forming.
n Clean the unit with a cloth and a little soapy water. Rinse with a damp cloth.
Then dry quickly with a cloth or pulsed air at 70°C max.
5.2 Metrological verification - Repair
n Like all measuring or testing devices, the instrument must be checked
regularly.
This instrument should be checked at least once a year. For checking and
calibration, contact one of our accredited metrology laboratories (information
and contact details available on request), at our Chauvin Arnoux subsidiary
or the branch in your country.
n Repair
For all repairs before or after expiry of warranty, please return the device to
your distributor.

Содержание