Accuracystatedfor5~100%ofrange;<20digitresidualreadingfor600mAopencircuit.
ACCrestFactormayreach3.0at4000counts;fornon‐sinusoidalwaveformstheCrestFactorerror
increasesperthefollowing:Add3%whenthecrestfactoris1~2;Add5%whenthecrestfactoris
2~2.5
Add7%whenthecrestfactoris2.5~3
Frequency(Hz) Resolution Accuracy OverloadProtect
10Hzto60kHz 0.001~0.01kHz ±(0.1%+4d) 600V
Notes:Inputamplitude≥10V(DClevel:0);Frequencyof65kHzorhigherforreferenceonly
Resistance Resolution Accuracy Overload
600.0Ω 0.1Ω
±(1.0%+2d) 600V
6.000kΩ 1Ω
±(0.8%+2d)
600V
60.00kΩ 10Ω
600V
600.0kΩ 100Ω
600V
6.000MΩ 1kΩ
±(1.2%+3d) 600V
60.00MΩ 10kΩ
±(1.5%+5d) 600V
Capacitance Resolution Accuracy Overload
6.200nF 1pF
InREL/ZEROmode:
±(4.0%+10d)
600V
62.00nF~620.0µF 10pF~0.1µF
±(4.0%+5d) 600V
6.200mF~62.00mF 1µF~10µF
±(10%) 600V
Note:Forbestaccuracy,useREL(MA61)orZEROfunction(MA63)whenmeasuring≤1µF
Continuity Resolution
Notes Overload
)))
0.1Ω
Beepersounds≤10Ω 600V
Diode Resolution
Notes Overload
1mV
Opencircuitvoltageis3.2V
approx.
Normalvoltagerange:
0.5V~0.8V
600V
Copyright©2015FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
Содержание
- True rms mini clamp meters 1
- Features 2
- Introduction 2
- International safety symbols 3
- Safety 3
- Safety notices 3
- 2 3 4 5 6 4
- Control buttons description 4
- Fig 1 meter parts 4
- Meter description 4
- Fig 2 function switch 5
- Function switch description 5
- Model ma61 5
- Model ma61 model ma63 5
- Model ma63 5
- Display icon descriptions 6
- Measurement unit prefixes 6
- Ω kω mω 6
- Automatic power off apo 7
- Current measurements clamping technique 7
- Fig 3 correct and incorrect clamping techniques 7
- Meter power 7
- Operation 7
- Ac current measurements 8
- Dc current measurements ma63 only 8
- Fig 4 correct and incorrect zero techniques 8
- Ac voltage measurements 9
- Fig 5 ac voltage measurement 9
- Frequency measurements 9
- Voltage measurements 9
- Dc voltage measurements 10
- Fig 6 dc voltage measurement 10
- Fig 7 resistance measurement 10
- Resistance measurements 10
- Capacitance measurements 11
- Continuity measurements 11
- Fig 8 capacitance measurement 11
- Diode test 12
- Fig 9 diode test 12
- Data hold 13
- Fig 10 non contact ac voltage detector 13
- Ncv non contact ac voltage detector 13
- Battery replacement 14
- Cleaning 14
- Zero relative measurements 14
- General specifications 15
- Specifications 15
- Range specifications 16
- Copyright 2015 flir systems inc 17
- Www extech com 17
Похожие устройства
- EXTECH PQ3210 Инструкция по эксплуатации
- EXTECH PQ3210 Инструкция EN
- EXTECH PQ3220 Инструкция по эксплуатации
- EXTECH PQ3220 Инструкция EN
- Fluke FLK-I17XX-FLEX3K-4 Инструкция по эксплуатации
- Fluke FLK-I17XX-FLEX3K-4 Инструкция EN
- Fluke FLK-323 Инструкция по эксплуатации
- Fluke FLK-323 Инструкция EN
- Fluke FLK-323 Документация
- Fluke FLK-323 Расширенная инструкция
- Fluke FLK-324 Инструкция по эксплуатации
- Fluke FLK-324 Инструкция EN
- Fluke FLK-324 Документация
- Fluke FLK-324 Расширенная инструкция
- Fluke FLK-325 Инструкция по эксплуатации
- Fluke FLK-325 Инструкция EN
- Fluke FLK-325 Документация
- Fluke FLK-325 Расширенная инструкция
- Fluke FLK-373 Инструкция по эксплуатации
- Fluke FLK-373 Инструкция EN