MeterDescription
123
4
5
678
9
1. EARTHterminalforInsulationResistanceandContinuitytesting.
2. GUARDterminalforACVandContinuitytesting.
3. VterminalforACVtestingandSelfTestmode.
4. LINETerminalforInsulationResistancetestingandSelfTestmode.
5. BacklitLCDDisplay
6. ZERObutton(shortpress)andLCDBacklight
ON/OFFbutton(longpress)
7. PI/DARTestingselectionbutton
8. TESTbutton
9. Rotaryfunctionselectswitch
Note:Securebatterycompartmentlocatedonrearofunit.
ControlButtons
1. PI/DARbutton:PresstoselectPIpolarizationindexandDARdielectricabsorptionratiotest
modes.
2. ZERO/Backlightbutton:Longpresstoenable/disableLCDbacklightorshortpresstoresetthe
displaytozeroforlowresistancemeasurements.
3. TESTbutton:Engage/disengagetestingforIR,Continuity,andACVoltagetests.
4. Rotaryfunctionswitch:SelectanIRoutputtestvoltage(250V/500V/1000V),measurelow
resistance(Continuity),ormeasureACVoltage(ACV).