5.8.Memory– cancellation of all measurements
Attention!
Before you delete all readings stored in the memory, make sure you
will not need refer to the readings at a later date.
In order to delete all stored readings, depress the MEMORY button for
at least 7 seconds, the display will show the symbol «CL» and 3 short
beep sounds will be heard to indicate deletion of stored readings.
Note:
If your last stored measurement was taken in the MAM mode, you
will see the « » symbol approximately 3 seconds during the
deletion procedure. (The display will show this symbol during
deletion of all stored measurements, if the last measurement is
taken in the MAM mode).
6. Error Messages/Malfunctions
If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(Example: Error no. 1)
* Please consult your doctor, if this or any other problem occurs repeatedly
M
1
Error No. Possible cause(s)
ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
This situation can occur, for example, if the tube becomes unplugged after the systolic blood
pressure has been measured. Further possible cause: No pulse has been detected.
ERR 2 Unnatural pressure impulses influence the measurement result. Reason: The arm was moved
during the measurement (Artefact).
ERR 3 The inflation of the cuff takes too long. The cuff is not correctly seated.
ERR 5 The measured results have indicated an unacceptable difference between the systolic and
diastolic pressure. Take another measurement, carefully following the instructions. Contact
your doctor, if you still obtain unusual results.
ERR 6 Single data differ too much during Average Mode even after 4 cycles. No average result can
be displayed.
HI The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats
per minute). Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40 beats per minute). Repeat the measurement.*