In case that the inflation pressure is not sufficient, the monitor
automatically re-inflates to a higher level.
c) When the instrument detects a pulse, the heart symbol in the
display starts to flash and a beep is heard for every heartbeat.
d) A longer beep is sounded when the measurement has been
completed. The systolic, and diastolic blood pressures and pulse
rate now appear in the display.
e) The measurement results are displayed, until you switch the
device off. If no button is pressed for 5 minutes, the device
switches automatically off, to save the batteries.
f) When the device is set to MAM mode, generally 3 separate
measurements will take place in succession and calculates your
detected blood pressure value. There will be 15 seconds resting time
in-between each measurement. A count down indicates the remaining
time and a beep will sound 5 seconds before the 2nd and 3rd readings
will begin. In case that the single data of each cycle differ too much
from each other, a fourth measurement is performed before the result
will be displayed. In rare cases the blood pressure is such unstable
that even after four measurements the data vary too much. In this
case «ERR 6» is shown and no result can be given.
If one of measurement causes an error message it is repeated.
5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any
reason (e.g. the patient feels unwell), the 0/I button can be pressed
at any time.
The device immediately lowers the cuff pressure automatically.
5.7. Memory – recall of measurements
This device automatically stores the last 30 measurements. By
pressing the MEMORY button, the number of stored measurements
is displayed (e.g. 30 set). By pressing again, the last measurement,
e.g. MR30 as well as previous 29 measurements (MR29, MR28,
…MR1) can be displayed one after the other. Data obtained in
MAM mode are indicated by a « » symbol.
MR30: Value of the last measurement – MR29: Value of the measurement before MR 30