6. Error messages/malfunctions
If an error occurs during a measurement, the measurement
is discontinued and a corresponding error code is displayed.
(Example Error no. 1)
* Please consult your doctor, if this or any other problem occurs repeatedly.
8
Error No. Possible cause(s)
ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
This situation can occur, for example, if the tube becomes unplugged after the systolic blood
pressure has been measured. Further possible cause: No pulse has been detected.
ERR 2 Unnatural pressure impulses influence the measurement result. Reason: The arm was moved
during the measurement (Artefact).
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated or the hose connection is
not sealed.
ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic
pressures. Take another reading following directions carefully. Contact you doctor if you
continue to get unusual readings.
HI The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats
per minute). Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40 beats per minute). Repeat the measurement.*
Malfunction Remedy
The display remains empty when the
instrument is switched on although the
batteries are in place.
1. Check whether the batteries are installed with the
correct polarity and correct if necessary.
2. If the display is unusual, remove the batteries and then
exchange them for new.
The pressure does not rise although the pump
is running.
Check the connection of the cuff tube and connect
properly if necessary
The device frequently fails to measure the
blood pressure values, or the values
measured are too low (too high).
1. Fit the cuff correctly on the arm.
2. Before starting the measurement make sure that the cuff
is not fitted too tightly and that there is no rolled-up
sleeve exerting pressure on the arm above the measuring
position. Take off articles of clothing if necessary.
3. Measure the blood pressure again in complete peace
and quiet.
Every measurement produces a different
value although the instrument functions
normally and the values displayed are nomal.
Please read the following information and the points
listed under «Common sources of error». Repeat the
measurement.

automatically 6 Error messages malfunctions If an error occurs during a measurement the measurement is discontinued and a corresponding error code is displayed Example Error no 1 Error No Possible cause s ERR1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg This situation can occur for example if the tube becomes unplugged after the systolic blood pressure has been measured Further possible cause No pulse has been detected ERR 2 Unnatural pressure impulses influence the measurement result Reason The arm was moved during the measurement Artefact ERR3 Inflation of the cuff takes too long The cuff is not correctly seated or the hose connection is not sealed ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures Take another reading following directions carefully Contact you doctor if you continue to get unusual readings HI The pressure in the cuff is too high over 300 mmHg OR the pulse is too high over 200 beats per minute Relax for 5 minutes and repeat the measurement LO The pulse is too low less than 40 beats per minute Repeat the measurement Please consult your doctor if this or any other problem occurs repeatedly Malfunction Remedy The display remains empty when the 1 Check whether the batteries are installed with the instrument is switched on although the correct polarity and correct if necessary batteries are in place 2 If the display is unusual remove the batteries and then The pressure does not rise although the pump Check the connection of the cuff tube and connect exchange them for new is running properly if necessary The device frequently fails to measure the 1 Fit the cuff correctly on the arm blood pressure values or the values 2 Before starting the measurement make sure that the cuff measured are too low too high is not fitted too tightly and that there is no rolled up sleeve exerting pressure on the arm above the measuring position Take off articles of clothing if necessary 3 Measure the blood pressure again in complete peace and quiet Every measurement produces a different Please read the following information and the points value although the instrument functions listed under Common sources of error Repeat the normally and the values displayed are nomal measurement 0