the arm. Any piece of clothing which does (e.g. a pullover)
must be taken off.
e) Secure the cuff with the Velcro closer in such a way that it
lies comfortably and is not too tight. Lay the arm on the
table (palm upwards) so that the cuff is at the same height
as the heart. Make sure that the tube is not kinked.
Note:
If it is not possible to fit the cuff to the left arm, it can also be
placed on the right one. However all measurements should be
made using the same arm.
5.4. Measuring procedure
After the cuff has been appropriately positioned, the measurement can begin:
a) Plug the T-connecting piece or the free end of the tube into
the left-hand side of the instrument.
b) Press the quick release valve at the pump ball for 3-4 sec.
to release any air out of the cuff.
c) Switch on the instrument by pressing the O/I button. The
tube should not yet be attached to the instrument. The
following appears in the display for 2 seconds: (Diag.). A
short be-be-beep follows to indicate that the instrument
has equilibrated with the pressure of the surroundings and
the measurement can therefore begin.
A «0» is shown in the display.
d) Grasp the pump ball with your free hand and inflate the cuff
to a pressure at least 40 mmHg higher than the expected
systolic pressure. If you do not know the value of this pres-
sure, inflate to a value of approximately 160–180 mmHg.
The cuff pressure is shown continuously on the LCD.
e) After the inflation process lay down the pump ball and remain
in a quiet and relaxed sitting position. The measurement now
proceeds on its own. If no measurement takes place and all
that appears in the display is a flashing, upwards pointing
arrow, see below under «Inflate further». When the
instrument detects a pulse, a heart symbol in the display
starts to flash and a beep is heard for every heart beat.
f) When the measurement has been concluded, a long peep tone
sounds. The measured systolic and diastolic blood pressure
values as well as the pulse frequency are now displayed.
The cuff pressure can now be quickly released by pressing
the release valve at the pump ball.
7
PULSE/min
SYS mmHg DIA mmHg