5.5. Inflating further
If the cuff is not inflated sufficiently, the measurement is broken off
after a few seconds and a flashing arrow pointing upwards appears
in the display. It is then necessary to inflate further to at least
20 mmHg higher than the previous value. The instruction to inflate
further can appear several times if the pressure is still not sufficient.
Attention:
If the cuff is inflated too far (over 300 mmHg), «HI» appears in
the display and a warning signal is sounded. The cuff pressure
must then be immediately released using the quick-release
valve (at the pump ball)! As soon as the cuff is inflated
sufficiently and the measurement is proceeding, continue as
described under points d–e.
5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for
any reason (e.g. the patient feels unwell), press the quick
release valve at the pump ball. The device then immediately
lowers the cuff-pressure automatically.
5.7. Memory – displaying the last measurement
The blood-pressure monitor automatically stores the last
measurement value. By pressing the MEMORY button, the
last measurement will be displayed.
6. Error messages/malfunctions
If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(example: error no. 3)
Error No. Possible cause(s)
ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
This situation can occur, for example, if the tube becomes unplugged after the systolic blood
pressure has been measured. Further possible cause: No pulse has been detected
ERR 2 The pressure dropped below 20 mmHg without a measurement taking place. The Release
valve was pressed during the measurement.
ERR 3 No systolic pressure could be detected. The cuff is not correctly seated.
ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic
pressures. Take another reading following directions carefully. Contact you doctor if you
continue to get unusual readings.
HI The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats
per minute). Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40 beats per minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs repeatedly.