Measuring body temperature on the forehead
Please note that the forehead/temples must be free from sweat and cosmetics and
that taking vasoconstrictive medication and skin irritations can distort the result
when measuring temperature on the forehead.
Press the • button for 1 second to switch the thermometer on. Following a
successful self test, the device emits two short beeps.
Place the measuring head with the forehead/cover cap fi tted on the temples, •
hold down the button and move the thermometer smoothly over the
forehead to the other temple.
Then release the • button. The end of the measuring time is signalled with a
short beep and the measured value appears on the display.
If the measured value is within the normal range (< 100.4°F/38°C), the green LED lights up for 3 seconds. If
the measured value is higher (≥ 100.4°F/38°C, high temperature), the red LED is illuminated.
Measuring surface temperature
Press the •
button for 1 second to switch the thermometer on. Following a successful self test, the
device emits two short beeps.
Then press the • and buttons simultaneously for 3 seconds to switch to the Object temperature
mode. The symbol appears in the display.
Press and hold the • or button and point the sensor tip towards to the object or liquid to be meas-
ured (never immerse in liquids), maintaining a distance of 3 cm.
Release the • or button. The end of the measuring time is signalled with a short beep and the
measured value appears on the display.
Please note that the displayed temperature is the measured and not the adapted surface temperature. It is
not the same as the forehead/ear temperature.
To switch back to the Ear/Forehead thermometer mode, press the and buttons simultaneously for 3
seconds until the symbol disappears and there is a short beep. By switching the thermometer o and on
again, it automatically exits the Object temperature mode.