ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacturer’s declaration – electromagnetic emissions
The BY 40 is intended for use in the electromagnetic environment specified below. The customer or the user of the
BY 40 should assure that it is used in such an environment
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The BY 40 uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11 Class B The BY 40 is suitable for use in all establishments Including domestic
and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The BY 40 is intended for use in the electromagnetic environment specified below. The customer or the user of the
BY 40 should assure that it is used in such an environment
Immunity test Test level Conformity level Electromagnetic environment –
guidelines
Electrostatic discharges
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %
Electrical fast transient/
burst
IEC 61000-4-4
Power supply lines ±2 kV
input/output lines ±1 kV
Power supply
lines ±2 kV
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Surge
IEC 61000-4-5
line(s) to line(s) ±1 kV.
line(s) to earth ±2 kV.
100 kHz repetition
frequency
line(s) to line(s) ±1 kV.
100 kHz repetition
frequency
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Voltage dips, short in
-
terruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycle
70% UT
(30% dip in UT)
for 25 cycle
<5% UT
(>95% dip in UT)
for 5s
5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycle
70% UT
(30% dip in UT)
for 25 cycle
<5% UT
(>95% dip in UT)
for 5s
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 Hz/60 Hz
30 A/m
50Hz/60Hz
Mains frequency magnetic fi eld
has to be at a level typical of a
commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage before to application of the test level.