Suunto Combo [10/12] Technical specifi cations
10
CAUTION!
Do not attempt to disassemble the gauges. Special tools and training are required to
perform the adequate service.
The instruments must be serviced by an authorized dealer every second year or after 200 dives
(whichever comes fi rst).
Operating and Storage Temperatures
SM-16 Operating temperature 0° C to 40° C [32° F to 104° F]
Storage temperature -20° C to 50° C [-4° F to 122° F]
SM-26 and SM-36 Operating and storage temperature 0° C to 40° C [32° F to 104° F]
Technical Specifi cations
The Suunto depth gauge SM-16 complies with the prEN 13319 standard.
The tolerances for depth gauges are:
At 3m +0.8 / -0.4 m
At 6m +0.8 / -0.4 m
At 9m ±0.8 m
At 15m +1/ -1.2 m
At 30m +1/ -1.2 m
At 45m +1/ -1.5 m
At 60m +1 / -1.5 m
The SM-26 and SM-36 pressure gauges comply with the EN 250 standard.
The tolerances for pressure gauges, measured at decreasing pressure are:
At 40 bar ± 5 bar
At 100 bar ± 10 bar
At 200 bar ± 10 bar
At 300 bar ± 10 bar