Fluke 117 [2/5] Change 1

Fluke 117 [2/5] Change 1
114, 115, 116, and 117 Calibration Information Manual Supplement
Change #1
On page 10, delete step 30.
Change #2, 543, 572
On page 4, replace Table 2 with the following:
Table 2. Accuracy Specifications
Function Range Resolution
Accuracy
± ([% of Reading] + [Counts])
Model
DC millivolts
600.0 mV
0.1 mV
0.5 % + 2
114, 115, 117
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 2 114, 115, 117
DC, 45 to 500
Hz
500 Hz to 1
kHz
Auto-V LoZ
[1]
True-
rms
600.0 V 0.1 V 2.0 % + 3 4.0 % + 3
114, 117
45 to 500 Hz
500 Hz to 1
kHz
AC millivolts
[1]
True-
rms
600.0 mV 0.1 mV
1.0 % + 3
2.0 % + 3
114, 115, 117
AC Volts
[1]
True-
rms
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3 2.0 % + 3 114, 115, 117
Continuity
600 1
Beeper on < 20 , off > 250 ;
detects opens or shorts of
500 µs or longer.
114, 115, 117
Ohms
600.0
6.000 k
60.00 k
600.0 k
6.000 M
40.00 M
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
5.0 % + 2
114, 115, 117
Diode test
2.000 V
0.001 V
0.9 % + 2
115, 117
Capacitance 1000 nF
10.00 µF
100.0 µF
9999 µF
1 nF
0.01 µF
0.1 µF
1 µF
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 µF - 1000 µF: 1.9 % + 2
> 1000 µF: 5 % + 20
115, 117
Lo-Z Capacitance
(Power-up option)
1 nF to 500 µF
10 % + 2 typical
115, 117
AC Amps True-
rms
[1]
(45 Hz to 500 Hz)
6.000 A
10.00 A
[3]
0.001 A
0.01 A
1.5 % + 3 115, 117
DC Amps
6.000 A
10.00 A
[3]
0.001 A
0.01 A
1.0 % + 3 115, 117
7/18 1
Скачать