Fluke 117 [4/5] Change 5
114, 115, 116, and 117 Calibration Information Manual Supplement
On page 3, under General Specifications, delete Surge Protection, Safety Compliance, and
Certifications and add:
Safety ..................................................... IEC 61010-1: Pollution Degree 2
IEC 61010-2-033: Measurement CAT III, CAT IV 600 V
Electromagnetic Compatibility (EMC)
International ....................................... IEC 61326-1: Portable Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio
frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and
those directly connected to a low-voltage power supply network that supplies buildings
used for domestic purposes. There may be potential difficulties in ensuring
electromagnetic compatibility in other environments due to conducted and radiated
disturbances.
Caution: This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is
connected to a test object.
Korea (KCC) ...................................... Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment
and the seller or user should take notice of it. This equipment is intended for use in
business environments and not to be used in homes.
USA (FCC) ........................................ 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Change #5, 593
On pages 11-12, in the Testing the VoltAlert Function (117 only) section, replace the procedure
and add Figure 1:
For Fluke 117 with S/N 38920290 through 42913406, use the alternate calibration procedure
described below to verify that VoltAlert functions properly.
Note
• Ensure the instrument is REMOVED from the holster prior to performing the test.
• Keep the meter away from electrical noise sources during the tests for, example, florescent
lights, dimmable lights, motors. These sources can trigger VoltAlert and invalidate the test.
• It may be necessary in steps 4 and 8 below to slightly adjust the Meter’s position for maximum
signal strength, in order to get the Meter’s beeper to sound continuously.
Refer to Figure 1 while performing the following steps.
1. Select the VoltAlert function, and verify that “” is on the display. Verify that the beeper is silent
and the red LED is off.
2. Connect wire (Pomona 4911A-12-2 or equivalent) to the High output voltage terminals of the
calibrator (Fluke 5500A or equivalent).
3. Set the calibrator output to 10 V at 60 Hz.
4. Hold the Meter so that the wire contacts and runs across the top center of the unit. The wire
should run perpendicular to the face of the unit (front to back of Meter, not side to side). Verify that
the Meter’s beeper is on continuously, and the red LED, at the top of the display, lights up.
5. Place the calibrator in standby mode and verify that the beeper is now silent and the red LED is
off.
6. Press the (display should indicate ‘’ range).
7. Set the calibrator’s output to 30 V at 60 Hz.
8. Hold the Meter so that the wire contacts and runs across the top center of the unit. The wire
should run perpendicular to the face of the unit (front to back of Meter, not side to side). Verify that
the Meter’s beeper is on continuously, and the red LED, at the top of the display, lights up.
9. Return the calibrator to standby mode and verify the Meter’s beeper is silent and the red LED is
off.
3 7/18
Содержание
- Manual supplement 1
- 115 116 and 117 calibration information manual supplement 2
- Change 1 2
- Change 2 2
- On page 10 delete step 30 2
- On page 4 replace table 2 with the following 2
- Table 2 accuracy specifications 2
- 63405 543 238 3
- Change 3 3
- Change 4 3
- On page 3 replace table 1 with 3
- Change 5 4
- Figure 1 voltalert alternate calibration procedure 5
Похожие устройства
- Fluke 117 Информация о калибровке
- Fluke 115 Положение об энергозависимости памяти
- Fluke 115 Дополнение к информации о калибровке
- Fluke 115 Информация о калибровке
- Fluke 114 Положение об энергозависимости памяти
- Fluke 114 Дополнение к информации о калибровке
- Fluke 114 Информация о калибровке
- Fluke 114 Extended Specs
- Fluke 115 Extended Specs
- Fluke 117 Extended Specs
- Fluke 975V AirMeter with Velocity Probe Положение об энергозависимости памяти
- Fluke 975V AirMeter with Velocity Probe Data Sheet
- Fluke 975V AirMeter with Velocity Probe Reference Guide
- Fluke 975V AirMeter with Velocity Probe Руководство по эксплуатации
- Fluke 424D Положение об энергозависимости памяти
- Fluke 424D Data Sheet
- Fluke 424D Краткий справочник
- Fluke 424D Паспорт безопасности
- Fluke 414D Краткий справочник
- Fluke 414D Паспорт безопасности