Fluke 714 [2/2] Change 1

Fluke 715 [2/2] Change 1
71X Series Calibration Manual Supplement
7/17 1
Change #1
On page 49, under mA Loop Power Verification, replace step 2 with:
2. Apply 1.2 k to the mA jacks from the decade box or other resistor source. The display must read
>19 mA but <O
L.
O
n page 54, under mA Loop Power Verification (718 Only), replace step 2 with:
2. Apply 1.2 k to the mA jacks from the decade box or other resistor source. The display must read
>19 mA but <O
L.
Change #2
On page 86, Table 71,
Change:
MP6
30G, 100G, 300G Pump Assembly
2558508
718
1
To:
MP6
Fluke-718, 30G/100G, Pump with clean out
2558508
718
1
MP6
Fluke-718, 300G, Pump with clean out
2632057
718
1
Change #3, 471
On page 34, replace Tables 47 and 48 with:
Table 47. 714 Thermocouple Measure Verification (mV)
Fluke 5520A
Fluke 714
-10.0000 mV
-10.012 mV to -9.988 mV
30.0000 mV
29.985 mV to 30.015 mV
75.0000 mV
74.979 mV to 75.021 mV
Table 48. 714 Thermocouple Source Verification (mV)
Fluke 714 Fluke 8508A
-10.00 mV -10.015 mV to -9.9885 mV
30.00 mV 29.9855 mV to 30.0145 mV
75.00 mV 74.97875 mV to 75.02125 mV