Fluke 712 [61/100] Pressure verification

Fluke 712 [61/100] Pressure verification
Process Calibrators
Verification
49
7. Disconnect the test leads and push to turn the power off.
mA Loop Power Verification
1. Hold down both LOOP POWER keys ( and ) at the same until the screen
reads:
---------- LOOP mA
then release the keys.
2. Apply 1 kΩ to the mA jacks from the decade box or other resistor sources. The
display should read over 19 mA but less than OL.
3. Push
to turn the power off.
Sensor Jack Verification
1. Make sure nothing is connected to the sensor port of the Fluke 717.
2. Turn power on. The display should read:
---------- mA
then change to:
0.000 PSI
3. Plug the Fluke Pressure Module into the sensor port on the Fluke 717. The display
should change to:
---------- mA
A pressure value should appear on the screen after all of the calibration constants
have been down loaded.
4. Disconnect the Fluke Pressure Module and push
to turn the Fluke 717 off.
718 and 718Ex Verification
This procedure is appropriate for the 718 and 718Ex. When the steps vary, the step will
be labeled (718 Only) or (718Ex Only).
Pressure Verification
WWarning
To avoid a violent release of pressure or vacuum, always
depressurize the system slowly using the pressure/vacuum
release control before detaching any pressure line.
1. Carefully attach the pressure fitting of the deadweight tester to the pressure port of
the Calibrator.
Note
The use of TEFLON tape at the pressure fitting strengthens the seal.
2. Push the green key to turn the calibrator on.
3. The display should read 0.00 PSI with the deadweight tester opened up to ambient
air. If not, push the
key until display does read 0.00 PSI.
4. Depending if the instrument being tested is a 718(Ex) 1G, 30G, 100G, or 300G, set
up the deadweight tester for the sequence of PSI inputs from Table 64 to be injected
into the pressure port of the Calibrator.
5. Ensure the pressure has stabilized at each input before verifying the display reading.
6. Carefully vent all pressure and disconnect the calibrator from the deadweight tester.