Fluke 724 [16/80] Temperature thermocouples 726

Fluke 726 [16/80] Temperature thermocouples 726
724/725/726
Calibration Manual
8
Temperature, Thermocouples (726)
Type
Minimum
(°C)
Maximum
(°C)
CJC ON
Accuracy
(°C)
CJC OFF
Accuracy
(°C)
J
-210 0.0 0.6 0.4
0.0 800 0.4 0.2
800 1200 0.5 0.3
K
-200 0.0 0.8 0.6
0.0 1000 0.5 0.3
1000 1372 0.7 0.5
T
-250 0.0 0.8 0.6
0.0 400 0.4 0.2
E
-250 -100 0.8 0.6
-100 1000 0.4 0.4
R
-20 0.0 2.0 1.8
0.0 1767 1.4 1.2
S
-20 0.0 2.0 1.8
0.0 1767 1.4 1.2
B
600 800 1.4 1.2
800 1000 1.5 1.3
1000 1820 1.7 1.5
C
0.0 1000 0.8 0.6
1000 2316 2.5 2.3
L
-200 0.0 0.45 0.25
0.0 900 0.4 0.2
U
-200 0.0 0.7 0.5
0.0 600 0.45 0.25
N
-200 0.0 1.0 0.8
0.0 1300 0.6 0.4
XK -200 800 0.4 0.2
BP
0.0 800 1.1 0.9
800 2500 2.3 2.1
Range Accuracy
Thermocouple in mV read -10 mV to 75 mV
0.015 % + 10 µV
(% of Reading +
Floor)
Thermocouple in mV source -10 mV to 75 mV
0.015 % + 10 µV
(% of Reading +
Floor)
Maximum current output in voltage ranges is 1 mA with an output impedance of 1 Ω
CJC error outside of 23 ± 5 °C is 0.05 °C / °C