Fluke 724 [31/80] Calibrators serial command list 726

Fluke 726 [31/80] Calibrators serial command list 726
Calibrators
Serial Command List (726)
23
Table 5. Serial Command List-726 and 725 V3.00 and Later (cont.)
Command Response/Actions
Command
Arguments
Comment Ch 1 Ch 2
CUST3_ALIAS?
(726 only)
Returns screen name for CUST3
RTD.
Verify RTD 3
alias
X
HART_ON Turns HART mode on. Switches in 250
Ω resistor
X X
HART_OFF Turns HART mode off. Switches out
250
Ω resistor
X X
HART? Returns state of hart mode, ON or
OFF
X X
*CLS Clear the error queue X X
FAULT Returns error code FILO X X
ERROR CODES:
NONNUMERIC_ENRTY (100)
EBUFFER_OVERFLOW (101)
INVALID_UNITS_CODE (102)
ENTRY_OVER_RNG (103)
ENTRY_UNDER_RNG (104)
MISSING_PARM (105)
INVALID_UNIT_PARM (106)
INVALID_SENSOR_TYPE (108)
UNKNOWN_COMMAND (110)
BAD_PARM_VALUE (111)
INPUT_BUFF_OVERFLOW (112)
MSG_BUFF_OVERFLOW (113)
OUTPUT_BUFF_OVERFLOW
(114)
OUTPUT_OVERLOAD (115)
X
X
CAL_START
Initiates a password protected
calibration (726 password = 627,
725 password = 527)
X X

Содержание