Fluke 724 [17/80] Temperature rtd ranges and accuracies 724 and 725

Fluke 726 [17/80] Temperature rtd ranges and accuracies 724 and 725
Calibrators
Specifications
9
Temperature, RTD Ranges, and Accuracies (724 and 725)
Type
Range
(°C)
Accuracy
Measure
4-wire (°C)
Measure*
2- and 3-wire (°C)
Source (°C)
Ni120 -80 to 260 0.2 0.3 0.2
Pt100-385 - 200 to 800 0.33 0.5 0.33
Pt100-392 -200 to 630 0.3 0.5 0.3
Pt100-JIS -200 to 630 0.3 0.5 0.3
Pt200-385
-200 to 250
250 to 630
0.2
0.8
0.3
1.6
0.2
0.8
Pt500-385
-200 to 500
500 to 630
0.3
0.4
0.6
0.9
0.3
0.4
Pt1000-385
-200 to 100
100 to 630
0.2
0.2
0.4
0.5
0.2
0.2
Resolution: 0.1 °C, 0.1 °F
Allowable excitation current (source): Ni120, Pt100-385, Pt100-392, Pt100-JIS,
Pt200-385: 0.05 to 0.80 mA,Pt500-385: 0.05 to 0.80 mA; Pt1000-385: 0.05 to 0.40 mA
RTD Source: Addresses pulsed transmitters and PLCs with pulses as short as 5 ms. SN < 7624001 may
need modification for pulses less than 15 ms.
*2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance not exceeding 100 Ω.