Fluke 724 [43/80] Lower display thermocouple measurement tests

Fluke 726 [43/80] Lower display thermocouple measurement tests
Calibrators
Performance Tests
35
Lower Display Thermocouple Measurement Tests
1. Remove the test leads from the UUT terminals; then connect a Type-J thermocouple
lead between the TC jack on the UUT and the TC jack on the 5520A, as shown in
Figure 10.
2. Press on the UUT until J appears on the lower display. If necessary, press
(use the configuration menu on the 726) so the temperature is displayed in °C.
3. Set the 5520A to output the Type-J thermocouple voltages shown in Table 13 and
verify the UUT temperature readings are within the limits shown (values use the ITS-
90 curves).
4. Press STBY on the 5520A.
aal11f.eps
Figure 10. Lower Display Thermocouple Test Connections
Table 13. Type-J Thermocouple Readings
5520A Settings
(referenced to 0 °C)
724/725 UUT Readings 726 UUT Readings (CJC On)
0.0 °C (0.000 mV) -0.7 °C to +0.7 °C -0.4 °C to +0.4 °C