Calibration Manual
14
Fuses
Warning
To avoid electrical shock:
Remove the test leads from the calibrator before opening the
battery door. Close and latch the battery door before using the
calibrator.
Use only the specified replacement fuses listed under
“Replaceable Parts”.
Do not allow water into the case.
Note
724 and 725 Version 2.x and lower.
Over time, the input protection has been modified to increase reliability. There will either
be self-resetting fuses, or replaceable fuses in sockets. The self-resetting fuses open
when heated by a current overload, and close when they cool down. When an input’s
fuse is open, the input’s functions will not work.
The time required for a self-resetting fuse to reset depends on the magnitude of the
overload. If a self-resetting fuse does not reset, return the Calibrator to an authorized
service center for repair.
If a replaceable fuse has been damaged, it needs to be replaced. The fuses can be
removed and checked for resistance. A value of < 10 is good.
The calibrator comes equipped with three 0.2 A 250 V socketed fuses.
Problems while measuring with the right jacks indicate that F3 may have opened.
Problems while measuring or sourcing with the center jacks or the TC jacks
indicate that F2 may have opened.
If you can’t measure or source current with the left jacks, F4 may have opened.
To access the fuses, refer to instructions under “Replacing the Batteries”. The fuse
compartment is located below the battery compartment. Use needle-nosed pliers to
remove them and test them with a multimeter. Replace the fuse with the proper
replacement fuse and follow the directions for reassembling the Calibrator. Refer to
Figure 1.