Fluke 725 [2/6] Change 1

Fluke 725 [2/6] Change 1
725 Users Manual Supplement
12/04 1
Change #1
On page 59, under RTD Excitation (simulation) change the following:
From:
PT 200-385 0.15 to 3.0 ma
To:
PT 200-385 0.05 to 0.80 mA
Under Temperature, RT Ranges, and Accuracies (ITS-90) replace the Allowable excitation
current (source) with the following:
Ni120, Pt100-385, Pt100-392, Pt100-JIS: 0.15 to 3.0 mA
Pt200-385, Pt500-385: 0.05 to 0.80 mA; Pt1000-385: 0.05 to 0.40 mA
Скачать