Tuttnauer 3870 EHS (HSG-WS) Инструкция по эксплуатации онлайн [41/58] 387790
![Tuttnauer 3870 EHS (HSG-WS) Инструкция по эксплуатации онлайн [41/58] 387790](/views2/1462512/page41/bg29.png)
39
12. At least once per week use a biological spore test (Bacillus
Stearothermophilus) in any load to insure proper sterilization. (Be aware
testing standard may vary).
13. Place instruments with ratchets opened and unlocked or clipped on the first
ratchet position.
14. Disassemble or sufficiently loosen multiple-part instruments prior to
packaging to permit the sterilizing agent to come into contact with all parts
of the instrument.
15. Tilt on edge items prone to entrap air and moisture, e.g. hollowware, so that
only minimal resistance to air removal, the steam passage and condensate
will be met.
16. Load items within the boundaries of the tray so that they do not touch the
chamber walls, or fall off when the tray is inserted into the autoclave.
17. The operator may use racks to allow for adequate separation of packaged
instruments.
18. Load trays loosely to capacity. Instruments should be loaded one level deep
only
19. Do not overload the sterilizer trays. Overloading will cause inadequate
sterilization and poor drying. maximum load is as follows:
Maximum solid load for model 2540 is 13. lbs and for model 3870 is 44 lbs.
Maximum textile load for model 2540 is 4.5. lbs and for model 3870 is
15.4 lbs.
20. Make sure that all instruments remain apart during the sterilization cycle.
21. Empty canisters should be placed upside-down, in order to prevent
accumulation of water.
22. Allow a distance of approximately 1” (2.5 cm) between trays or cassettes to
permit steam circulation.
23. Wrapped instruments should be packed in material that will allow steam
penetration and promote drying, such as autoclave bag, autoclave paper, or
muslin towels.
24. Do not stack pouches. It is recommended that a Tuttnauer Pouch Rack be
used. This will allow the operator to place pouches on their side, which will
increase capacity and provide proper spacing to ensure steam penetration
and promote adequate drying.
Содержание
- Electronic table top pre and post vacuum autoclave 1
- Maintenance manual 1
- Models 2540 3870 ehs 1
- Operation 1
- General 6
- General information 8
- Sterilization programs 21
- A second test is done after 10 minutes to establish p3 and to determine the vacuum lost p3 p2 27
- Note during the test period the autoclave is not heated note during the test period the autoclave is not heated 27
- Operations sequence operations sequence vacuum is produced in the chamber down to 12 psig 25 ih 27
- Program 6 6 vacuum leakage test vacuum is produced in the chamber down to p1 12 psig 25 ih at this stage all the valves close the autoclave remains in this stage for 5 minutes this period enables the condition in the chamber to reach equilibrium after the 5 minutes have elapsed the printer records the pressure that is referred to as p2 at this point the test begins and lasts 10 minutes at the end of the test the printer records the results the pressure at the end of the test is referred to as p3 the rate of change of p3 to p2 shall not exceed 0 2 psi min if p3 to p2 exceeds 0 2 psi min the printer will print fail if p3 to p2 is within the required value the cycle will end normally 27
- The chamber is allowed to stabilize for 5 minutes then a test is done to establish p2 27
- The vacuum pump stops 27
- Front panel keyboard 28
- Keyboard keys and display 28
- Flash 273 st by 230 º f j30 14 33
- Wdry 250 exh low temp 25 33
- Wdry 273 ster 273 º f 03 14 30 33
- Printer 36
- Preparation before sterilization 40
- Dental handpiece sterilization 43
- Operation 45
- Maintenance instructions 48
- Troubleshooting for the operator 55
- Accessories 58
- Spare parts list 58
Похожие устройства
- Tuttnauer 3870 E Инструкция по эксплуатации
- Tuttnauer 3850 M Инструкция по эксплуатации
- Woson TANZO C23 NEW Инструкция по эксплуатации
- Woson TANZO C12 Инструкция по эксплуатации
- Woson TANZO C18 NEW Инструкция по эксплуатации
- Woson TANZO C23 Инструкция по эксплуатации
- Woson TANZO C12 NEW Инструкция по эксплуатации
- Woson TANZO E23 touch Инструкция по эксплуатации
- Woson TANZO C18 Инструкция по эксплуатации
- Woson TANZO Е18 Инструкция по эксплуатации
- Woson TANZO E18 touch Инструкция по эксплуатации
- Youjoy BES-12L-B Инструкция по эксплуатации
- Youjoy BES-17L-B Инструкция по эксплуатации
- Youjoy BES-15L-B Инструкция по эксплуатации
- Youjoy BES-22L-B Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-20 СПУ Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-80 СПУ Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-40 СПУ Инструкция по эксплуатации
- J.Morita HIM-1 Lubrina Инструкция по эксплуатации
- Nsk Care3 Plus C1 Инструкция по эксплуатации