Tuttnauer 3870 EHS (HSG-WS) Инструкция по эксплуатации онлайн [42/58] 387790

Tuttnauer 3870 EHS (HSG-WS) Инструкция по эксплуатации онлайн [42/58] 387790
40
25. Packs
1. Place packs upright on trays, side by side.
2. Packs should not touch the chamber walls.
3. Pack instrument sets in a manner that prevents damage to delicate items.
4. Pack hollowware sets so that all openings face the same direction and so
that the contents cannot move inside the pack.
5. Load packs of folded operating room drapes with layers vertical,
allowing air to be removed from the packs rapidly.
6. Do not place packs of hollowware and trays of instruments above textile
packs or soft goods in order to avoid wetting caused by condensation
from items above.
7. Load items packed in flexible packaging materials on edge with paper to
laminate, or flat with the plastic surface downwards.
Note:
The manufacturer’s recommendations shall be observed,
concerning the sterilization data for each type of material.
26. Tubing
When placing in a tray, make sure that both ends are open, without sharp
bends or twists.
27. Cassettes
Instruments may be sterilized in cassettes. The advantage of the cassettes is
that the sterilized instruments may remain organized in the cassettes ready
for use, while stored in a sterile area.
If using model 2540 remove the trays and slide the cassettes into the
chamber on the rack system.
If using model 3870 place the cassettes directly on the tray, either lying flat
(but no stacking) or on edge.
Wrong
Right