Tuttnauer 3870 EHS (HSG-WS) [40/58] Preparation before sterilization
![Tuttnauer 3870 EHS (HSG-WS) [40/58] Preparation before sterilization](/views2/1462512/page40/bg28.png)
38
6 PREPARATION BEFORE STERILIZATION
This section applies to the sterilization of all instruments [except for dental
handpieces]. For the preparation of dental handpieces, please refer to the
handpiece manufacture’s instructions and see section 7 in this manual.
CAUTION: The liquid cycle is not meant for the sterilization of liquids
intended for direct patient contact.
The purpose of packaging and wrapping items for sterilization is to provide an
effective barrier against contamination during storage, once the items have been
sterilized.
Packaging and wrapping materials should be approved for use in a steam
sterilizer and permit the removal of air and penetration of the steam during the
sterilization process.
The basic principle of determining the size, mass and contents of instrument and
hollowware packs is that the contents are sterile and dry immediately upon
completion of the drying cycle
Instruments to be sterilized must be clean and free from any residual matter,
such as debris, blood, pads or any other material. Such substances may cause
damage to the instruments themselves or the sterilizer.
1. Clean instruments immediately after use. It is recommended that instruments
be ultrasonically cleaned using Tuttnauer’s Clean and Simple enzymatic
cleaning solution.
2. After ultrasonic cleaning rinse under tap water for 30 seconds and pat dry to
remove residual minerals. If your tap water has a high mineral content then
rinse a second time in a bath of distilled water or alcohol to remove minerals.
3. Launder textile wraps prior to reuse, but do not use bleach.
4. Follow the instrument manufacturer’s instructions for cleaning and
lubricating instruments.
5. Be sure that instruments of dissimilar metal (stainless steel, carbon steel,
etc.) are separated. Carbon steel instruments should be bagged or placed on
autoclavable towels and not directly on stainless steel trays. (Mixing will
result in the oxidation of these metals).
6. Do not place materials to be sterilized directly on the chamber’s wall. Place
the material only on trays, rack, etc.
7. Check the manufacture’s instructions as to the proper procedure for
sterilizing each item.
8. Place a sterilization indicator in each tray or inside each wrapped pack.
9. All instruments must be sterilized in an open position.
10. Use single-use wraps once only and discard after use.
11. Verify that the packaging method is in accordance with good practice
approach and the packaging materials are in accordance with the applicable
standards (e.g. EN868 series).
Содержание
- Electronic table top pre and post vacuum autoclave 1
- Maintenance manual 1
- Models 2540 3870 ehs 1
- Operation 1
- General 6
- General information 8
- Sterilization programs 21
- A second test is done after 10 minutes to establish p3 and to determine the vacuum lost p3 p2 27
- Note during the test period the autoclave is not heated note during the test period the autoclave is not heated 27
- Operations sequence operations sequence vacuum is produced in the chamber down to 12 psig 25 ih 27
- Program 6 6 vacuum leakage test vacuum is produced in the chamber down to p1 12 psig 25 ih at this stage all the valves close the autoclave remains in this stage for 5 minutes this period enables the condition in the chamber to reach equilibrium after the 5 minutes have elapsed the printer records the pressure that is referred to as p2 at this point the test begins and lasts 10 minutes at the end of the test the printer records the results the pressure at the end of the test is referred to as p3 the rate of change of p3 to p2 shall not exceed 0 2 psi min if p3 to p2 exceeds 0 2 psi min the printer will print fail if p3 to p2 is within the required value the cycle will end normally 27
- The chamber is allowed to stabilize for 5 minutes then a test is done to establish p2 27
- The vacuum pump stops 27
- Front panel keyboard 28
- Keyboard keys and display 28
- Flash 273 st by 230 º f j30 14 33
- Wdry 250 exh low temp 25 33
- Wdry 273 ster 273 º f 03 14 30 33
- Printer 36
- Preparation before sterilization 40
- Dental handpiece sterilization 43
- Operation 45
- Maintenance instructions 48
- Troubleshooting for the operator 55
- Accessories 58
- Spare parts list 58
Похожие устройства
- Tuttnauer 3870 E Инструкция по эксплуатации
- Tuttnauer 3850 M Инструкция по эксплуатации
- Woson TANZO C23 NEW Инструкция по эксплуатации
- Woson TANZO C12 Инструкция по эксплуатации
- Woson TANZO C18 NEW Инструкция по эксплуатации
- Woson TANZO C23 Инструкция по эксплуатации
- Woson TANZO C12 NEW Инструкция по эксплуатации
- Woson TANZO E23 touch Инструкция по эксплуатации
- Woson TANZO C18 Инструкция по эксплуатации
- Woson TANZO Е18 Инструкция по эксплуатации
- Woson TANZO E18 touch Инструкция по эксплуатации
- Youjoy BES-12L-B Инструкция по эксплуатации
- Youjoy BES-17L-B Инструкция по эксплуатации
- Youjoy BES-15L-B Инструкция по эксплуатации
- Youjoy BES-22L-B Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-20 СПУ Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-80 СПУ Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-40 СПУ Инструкция по эксплуатации
- J.Morita HIM-1 Lubrina Инструкция по эксплуатации
- Nsk Care3 Plus C1 Инструкция по эксплуатации