2 GENERAL INFORMATION
2.1 Introduction
The EHS is a high speed, high vacuum, high capacity, hospital grade
sterilizer with FDA cleared flash cycle. A built in vacuum pump and
steam generator allows for increased load size and dramatically reduces
total cycle time. The closed door drying system provides a faster and
more thorough drying cycle. Six automatic cycles are available. Four
sterilization cycles and two test cycles. EHS models are available in a
10” or 15” chamber.
The EHS autoclave, models 2540 and 3870 are designed for the
sterilization of wrapped and unwrapped instruments and related items
found in dental, medical and veterinary clinics, first aid rooms,
hospitals, laboratories etc. They are electrically – heated pre and post
vacuum sterilizers, using steam as the sterilizing agent. A computerized
control unit ensures a fully automatic sterilization cycle, and the control
and monitoring of physical parameters.
The EHS pre / post vacuum sterilizer has the following features:
¾ It is a Class B sterilizer in accordance with the European standard
prEN 13060-2
¾ It has a pre-vacuum conditioning stage
¾ It has a post-vacuum drying phase.
¾ It has a jacketed chamber
¾ It has a dual compartment water reservoir
¾ It has a complete water discharge system
¾ It has a PC friendly communication system
The advantages of this pre / post vacuum sterilizer are:
¾ Dramatically faster overall cycle time
¾ Better and deeper steam penetration into loads
¾ Increased load size
¾ More thorough and efficient drying of materials
¾ Automatically maintained water reservoirs
¾ Immediate availability of steam for sterilization
¾ Non recycled water reduces maintenance
The Vacuum pump:
The pre and post vacuum stages are achieved with the use of a high
volume liquid ring vacuum pump.
The advantage of the pre-vacuum stage is felt during the sterilization
stage. Vacuum removal of at least 99.5% of the air in the chamber
allows the steam to penetrate deeper into packed materials. Deeper
penetration allows for larger loads to be sterilized, thus saving time by
having to run fewer cycles.
The advantage of the post-vacuum stage is an efficient and thorough
drying of bagged and wrapped items. This is accomplished by the
combined use of a heated chamber and the vacuum removal of residual
moisture.
Содержание
- Electronic table top pre and post vacuum autoclave 1
- Maintenance manual 1
- Models 2540 3870 ehs 1
- Operation 1
- General 6
- General information 8
- Sterilization programs 21
- A second test is done after 10 minutes to establish p3 and to determine the vacuum lost p3 p2 27
- Note during the test period the autoclave is not heated note during the test period the autoclave is not heated 27
- Operations sequence operations sequence vacuum is produced in the chamber down to 12 psig 25 ih 27
- Program 6 6 vacuum leakage test vacuum is produced in the chamber down to p1 12 psig 25 ih at this stage all the valves close the autoclave remains in this stage for 5 minutes this period enables the condition in the chamber to reach equilibrium after the 5 minutes have elapsed the printer records the pressure that is referred to as p2 at this point the test begins and lasts 10 minutes at the end of the test the printer records the results the pressure at the end of the test is referred to as p3 the rate of change of p3 to p2 shall not exceed 0 2 psi min if p3 to p2 exceeds 0 2 psi min the printer will print fail if p3 to p2 is within the required value the cycle will end normally 27
- The chamber is allowed to stabilize for 5 minutes then a test is done to establish p2 27
- The vacuum pump stops 27
- Front panel keyboard 28
- Keyboard keys and display 28
- Flash 273 st by 230 º f j30 14 33
- Wdry 250 exh low temp 25 33
- Wdry 273 ster 273 º f 03 14 30 33
- Printer 36
- Preparation before sterilization 40
- Dental handpiece sterilization 43
- Operation 45
- Maintenance instructions 48
- Troubleshooting for the operator 55
- Accessories 58
- Spare parts list 58
Похожие устройства
- Tuttnauer 3870 E Инструкция по эксплуатации
- Tuttnauer 3850 M Инструкция по эксплуатации
- Woson TANZO C23 NEW Инструкция по эксплуатации
- Woson TANZO C12 Инструкция по эксплуатации
- Woson TANZO C18 NEW Инструкция по эксплуатации
- Woson TANZO C23 Инструкция по эксплуатации
- Woson TANZO C12 NEW Инструкция по эксплуатации
- Woson TANZO E23 touch Инструкция по эксплуатации
- Woson TANZO C18 Инструкция по эксплуатации
- Woson TANZO Е18 Инструкция по эксплуатации
- Woson TANZO E18 touch Инструкция по эксплуатации
- Youjoy BES-12L-B Инструкция по эксплуатации
- Youjoy BES-17L-B Инструкция по эксплуатации
- Youjoy BES-15L-B Инструкция по эксплуатации
- Youjoy BES-22L-B Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-20 СПУ Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-80 СПУ Инструкция по эксплуатации
- Смоленское Сктб Спу ГП-40 СПУ Инструкция по эксплуатации
- J.Morita HIM-1 Lubrina Инструкция по эксплуатации
- Nsk Care3 Plus C1 Инструкция по эксплуатации