Fluke 27 [14/30] Resistance and diode test continuity
![Fluke 27 [14/30] Resistance and diode test continuity](/views2/1005102/page14/bge.png)
27
Users Manual
10
Resistance and Diode Test/Continuity
Caution
To avoid possible damage to the meter or to
the equipment under test, disconnect circuit
power and discharge all high voltage
capacitors before measuring resistance,
testing for continuity, or diode test.
Resistance Measurement
Test lead resistance may influence measurement
accuracy on the 320-ohm range. To determine the error,
short the test leads together and read the lead
resistance. Correct the measurement by subtracting the
lead resistance from the measurement, or use the
Relative (REL) mode to zero the display. The error is
usually 0.1 to 0.2 ohms for a standard pair of test leads.
Some in-circuit resistance measurements can be made
without removing diodes and transistors from the circuit.
The full-scale measurement voltage produced on ranges
below 32 megohms does not strongly forward bias silicon
diodes or transistor junctions. Use the highest range you
can (except 32 megohm) to minimize the possibility of
turning on diodes or transistor junctions. Full scale
measurement voltage in the 32-megohm range does
strongly forward bias a diode or transistor.
Diode Test and Continuity
In diode test, there is only one range: 0 to +2.08 volts.
Voltage is developed across the component(s) under test
by a test current output from the Fluke 27. Voltages
greater than 2.08 V or open test leads produce an
overload (OL) condition. Negative inputs produce a
negative indication (they are not suppressed). In the
diode test function S the beeper produces a
continuous tone if the input is less than 0.1 V, and the
beeper beeps once when the input descends through a
0.7 V threshold.
Audible continuity testing is also performed with the
function selector switch in the diode test/continuity
position. A continuous tone sounds for test resistances
below approximately 270 ohms. An intermittent
connection produces erratic beeps, and can be valuable
troubleshooting aid. Erratic beeps can also occur, due to
environmental noise, if a test value is very close to the
threshold (270 ohms). Test resistances from
approximately 270 ohms to 2600 ohms produce a short
tone similar to a forward biased diode. Test resistances
less than approximately 36 kilohms will produce an on-
scale reading.
Содержание
- For iec 61010 cat iii meters only 1
- Multimeter 1
- Fluke corporation fluke europe b v p o box 9090 p o box 1186 everett wa 5602 b d eindhoven 98206 9090 the netherlands 2
- Limited warranty limitation of liability 2
- Register your meter at register fluke com 2 02 2
- Visit the fluke website at www fluke com 2
- Table of contents 3
- Title page 3
- Users manual 4
- Multimeter safety 5
- Warnings 5
- A amperes input terminal 9
- Com common terminal 9
- Digital display 9
- Function selector rotary switch 9
- Milliamp microamp input terminal 9
- Operating features 9
- Volt ohms diode test input terminal 9
- Min max mode pushbutton 10
- Range manual range mode pushbutton 10
- Rel relative mode pushbutton 10
- Warning 10
- Hold touch hol 11
- Max maximum annunciator 11
- Min minimum annunciator 11
- Mk resistance annunciators 11
- Mode annunciator 11
- Mode pushbutton 11
- Ns conductance range annunciator ns 11
- Relative annunciator 11
- Touch hol 11
- Analog bar graph display 12
- Decimal point range indicator 12
- Low battery annunciator 12
- Manual range annunciator 12
- Negative polarity annunciator 12
- Overflow indication 12
- Overload indication 12
- Warning 12
- Applications 13
- Beeper not illustrated 13
- Current ac dc 13
- Tilt bail adjustment 13
- Voltage ac dc 13
- Warning 13
- Diode test and continuity 14
- Resistance and diode test continuity 14
- Resistance measurement 14
- Conductance 15
- Leakage testing 15
- Analog bar graph application 16
- Using the analog bar graph 16
- Specific applications checking capacitors 17
- Specific applications contact bounce 17
- Specific applications nulling 17
- Operator maintenance 20
- Specific applications noisy resistance measurements 20
- Warnings 20
- Battery installation or replacement 21
- Figure 1 battery and fuse replacement 22
- Users manual 22
- Fuse replacement 23
- Fuse test 23
- Warning 23
- Warnings 23
- General maintenance 24
- Figure 2 replaceable parts 25
- Operator maintenance 25
- Service 26
- Specifications 27
Похожие устройства
- LG MS-2352T Инструкция по эксплуатации
- Ada 6D Maxliner А00138 Инструкция по эксплуатации
- Supra SFD-115U Инструкция по эксплуатации
- DBX DRIVERACK PA+ 2X6 Инструкция по эксплуатации
- Belkin AV10009qp2M Инструкция по эксплуатации
- LG E1069SD Инструкция по эксплуатации
- Supra SFD-111U Инструкция по эксплуатации
- Blue Weld Combi 132 821340 Инструкция по эксплуатации
- Fluke 27II Инструкция по эксплуатации
- LG MS-2352JS Инструкция по эксплуатации
- DBX DRIVERACK PA-EU Инструкция по эксплуатации
- LG F1480TDS5 Инструкция по эксплуатации
- Supra SFD-105U Инструкция по эксплуатации
- Geo-Fennel FL 40-Pocket II-HP 541100 Инструкция по эксплуатации
- LG MS-2352J Инструкция по эксплуатации
- DBX DRIVERACK PX Инструкция по эксплуатации
- Belkin F5C412eb300W Инструкция по эксплуатации
- Fluke 28II Инструкция по эксплуатации
- LG F1222NDR5 (Серебристый) Инструкция по эксплуатации
- Supra SFD-103U Инструкция по эксплуатации