Analog Bar Graph Application
13
graph excels at displaying trends, or slowly changing
signals. In addition, autoranging on the Fluke 27 allows
monitoring the signal change through changing ranges.
Many diagnostic routines using the bar graph require
practice. The operator is looking for good or bad signal
patterns that occur over some span of time. Capacitance
checks and noisy resistance measurements create such
patterns. Therefore, familiarity with analog bar graph
response and movement is necessary to accurately
interpret a signal pattern. Compare the bar graph
response when making measurements on a known-good
unit to the bar graph response when making
measurements on a faulty unit.
Specific Applications--Nulling
The Fluke 27 bar graph is ideal for nulling adjustments.
As an adjustment approaches zero, fewer bar graph
segments are displayed, then no bar graph segments are
displayed. The - annunciator flickers when the input level
is within 10 counts to zero. The flickering null indication is
displayed every time the input approaches zero or swings
from one polarity to the other. The operator merely
watches for the - annunicator indication, then reverses
the direction of the adjustment when the polarity sign is
displayed. In one or two passes, a near-zero input level is
possible, then the digital display can be used for exact
zero adjustment.
Specific Applications--Contact Bounce
When subject to vibration, relay contacts may begin to
bounce open. Checking for this intermittent problem is a
routine troubleshooting measure associated with many
types of equipment, including computers. Since the
bounce problem will worsen as the relay fatigues, early
diagnosis is important.
When the contact bounces open, its resistance value
changes momentarily from zero to infinity and back.
Ordinary hand-held DMMs take more than 300
milliseconds to update their displays--much too long to
detect a brief contact bounce. A traditional VOM needle
will move slightly at the instant of contact bounce, but the
inertia of the needle movement dampens the response.
The analog bar graph, however, will display at least one
segment the moment the contact opens. The bar graph
can detect contact bounce as brief as 0.2 milliseconds,
while most analog needle movements require a 3
millisecond opening before they will respond.
Specific Applications--Checking Capacitors
Volt-ohm meters are often used as simple capacitor
checkers. In the capacitor kick test, the needle of the
VOM in the resistance mode moves quickly from open
(infinite ohms) toward short (zero ohms) as the capacitor
is placed across the VOM input. The VOM battery
Содержание
- For iec 61010 cat iii meters only 1
- Multimeter 1
- Fluke corporation fluke europe b v p o box 9090 p o box 1186 everett wa 5602 b d eindhoven 98206 9090 the netherlands 2
- Limited warranty limitation of liability 2
- Register your meter at register fluke com 2 02 2
- Visit the fluke website at www fluke com 2
- Table of contents 3
- Title page 3
- Users manual 4
- Multimeter safety 5
- Warnings 5
- A amperes input terminal 9
- Com common terminal 9
- Digital display 9
- Function selector rotary switch 9
- Milliamp microamp input terminal 9
- Operating features 9
- Volt ohms diode test input terminal 9
- Min max mode pushbutton 10
- Range manual range mode pushbutton 10
- Rel relative mode pushbutton 10
- Warning 10
- Hold touch hol 11
- Max maximum annunciator 11
- Min minimum annunciator 11
- Mk resistance annunciators 11
- Mode annunciator 11
- Mode pushbutton 11
- Ns conductance range annunciator ns 11
- Relative annunciator 11
- Touch hol 11
- Analog bar graph display 12
- Decimal point range indicator 12
- Low battery annunciator 12
- Manual range annunciator 12
- Negative polarity annunciator 12
- Overflow indication 12
- Overload indication 12
- Warning 12
- Applications 13
- Beeper not illustrated 13
- Current ac dc 13
- Tilt bail adjustment 13
- Voltage ac dc 13
- Warning 13
- Diode test and continuity 14
- Resistance and diode test continuity 14
- Resistance measurement 14
- Conductance 15
- Leakage testing 15
- Analog bar graph application 16
- Using the analog bar graph 16
- Specific applications checking capacitors 17
- Specific applications contact bounce 17
- Specific applications nulling 17
- Operator maintenance 20
- Specific applications noisy resistance measurements 20
- Warnings 20
- Battery installation or replacement 21
- Figure 1 battery and fuse replacement 22
- Users manual 22
- Fuse replacement 23
- Fuse test 23
- Warning 23
- Warnings 23
- General maintenance 24
- Figure 2 replaceable parts 25
- Operator maintenance 25
- Service 26
- Specifications 27
Похожие устройства
- LG MS-2352T Инструкция по эксплуатации
- Ada 6D Maxliner А00138 Инструкция по эксплуатации
- Supra SFD-115U Инструкция по эксплуатации
- DBX DRIVERACK PA+ 2X6 Инструкция по эксплуатации
- Belkin AV10009qp2M Инструкция по эксплуатации
- LG E1069SD Инструкция по эксплуатации
- Supra SFD-111U Инструкция по эксплуатации
- Blue Weld Combi 132 821340 Инструкция по эксплуатации
- Fluke 27II Инструкция по эксплуатации
- LG MS-2352JS Инструкция по эксплуатации
- DBX DRIVERACK PA-EU Инструкция по эксплуатации
- LG F1480TDS5 Инструкция по эксплуатации
- Supra SFD-105U Инструкция по эксплуатации
- Geo-Fennel FL 40-Pocket II-HP 541100 Инструкция по эксплуатации
- LG MS-2352J Инструкция по эксплуатации
- DBX DRIVERACK PX Инструкция по эксплуатации
- Belkin F5C412eb300W Инструкция по эксплуатации
- Fluke 28II Инструкция по эксплуатации
- LG F1222NDR5 (Серебристый) Инструкция по эксплуатации
- Supra SFD-103U Инструкция по эксплуатации