Fluke 27 [27/30] Specifications

Fluke 27 [27/30] Specifications
Specifications
23
Specifications
Function Range Resolution Accuracy*
3.200 V 0.001 V ±(0.1%+1)
C 32.00 V 0.01 V ±(0.1%+1)
320.0 V 0.1 V ±(0.1%+1)
1000 V 1 V ±(0.1%+1)
J
320.0 mV 0.1 mV ±(0.1%+1)
320.0 e 0.1 e ±(0.3%+3)**
3.200 ke 0.001 ke ±(0.2% +1)
O 32.00 ke 0.01 ke ±(0.2% +1)
(nS) 320.0 ke 0.1 ke ±(0.2% +1)
3.200 Me 0.001 Me ±(0.2% +1)
32.00 Me 0.01 Me ±(1%+1)
32.00 nS 0.01 nS ±(2%+10)
S 2.080 V 0.001 V ±(1%+1) typical
40 Hz-2 kHz 2 kHz-10 kHz 10 kHz-30 kHz
3.200 V 0.001 V ±(0.5%+3) ±(2%+3) ±(4%+10)
I 32.00 V 0.01 V ±(0.5%+3) ±(2%+3) ±(4%+10)
320.0 V 0.1 V ±(0.5%+3) ±(2%+3) ±(4%+10)
1000 V 1 V ±(1%+3) ±(3%+3) Not Specified
F 320.0 mV 0.1 mV ±(0.5%+3) ±(2%+3) ±(4%+10)
** When using the REL () function to compensate for offsets

Содержание