Fluke 53 II [16/26] Enabling or disabling sleep mode
![Fluke 53 II [16/26] Enabling or disabling sleep mode](/views2/1005327/page16/bg10.png)
53 & 54 Series II
Users Manual
12
Enabling or Disabling Sleep Mode
The thermometer enters sleep mode if no button press
occurs for 20 minutes.
Pressing any button wakes the thermometer and returns it
to its previous state.
1. Press
K or N until the display shows SLP.
2. Press
E to indicate that you want to change the
sleep setting.
The display shows on if sleep mode is on and 0FF if
sleep mode is off.
3. Press
K or N as needed until the display shows
on or 0FF.
4. Press
E to store the sleep setting in memory.
Sleep mode is enabled each time you turn on the
thermometer and is automatically disabled in MIN MAX
and logging modes.
Setting the Time
1. Press K or N until the display shows the time if it
is set or shows "- - : - -."
2. Press
E to indicate you want to set the time.
The left two digits blink.
3. Press
K or N until the display shows the correct
hour (24-hour format), and then press
E to select.
The right two digits blink.
4. Press
K or N until the display shows the correct
minutes, and then press
E to store the time in
memory.
Note
Holding down
K
or
N
causes the number to
change more quickly.
Содержание
- 54 series ii 1
- 54 series ii thermometer 1
- September 1999 rev 6 01 1
- Thermometer 1
- Users manual 1
- Limited warranty limitation of liability 2
- Table of contents 3
- Title page 3
- 54 series ii 4
- 54 series ii 5
- Contacting fluke 5
- Safety information 5
- Warning 6
- Warning cont 7
- Getting started 8
- Components 9
- Display elements 10
- Buttons 11
- Changing setup options 13
- Entering and exiting setup 13
- Using the thermometer 13
- Changing the logging interval 14
- Changing the offset 15
- Changing the thermocouple type 15
- Enabling or disabling sleep mode 16
- Setting the time 16
- Changing the line frequency 17
- Connecting a thermocouple 17
- Measuring temperatures 17
- Displaying temperatures 18
- Holding the displayed temperatures 18
- Viewing the min max and avg readings 18
- Using memory 19
- Using the offset to adjust for probe errors 19
- Initial conditions and data entries 20
- Starting and stopping logging 20
- Clearing memory 21
- Viewing logged readings 21
- Communicating with a pc 22
- Calibration 23
- Cleaning the case and holster 23
- Environmental 23
- Maintenance 23
- Replacing the batteries 23
- Specifications 23
- Electrical 24
- General 24
- Pk 1 thermocouple supplied with thermometer 24
- Electrical cont 25
- Replacement parts and accessories 25
Похожие устройства
- Indesit WISE 10 CSI Инструкция по эксплуатации
- LG MS-1744U Инструкция по эксплуатации
- Bosch PHC 9690 ProSalon Инструкция по эксплуатации
- Geo-Fennel Square Linner II 551000 Инструкция по эксплуатации
- Supra SWM-755 Инструкция по эксплуатации
- LG MS-1744HL Инструкция по эксплуатации
- Indesit IWSE 5105 Инструкция по эксплуатации
- Fluke 54 II Инструкция по эксплуатации
- Geo-Fennel Ecoline EL515 D1000 Инструкция по эксплуатации
- Fluke 561 Инструкция по эксплуатации
- LG MS-1724WB Инструкция по эксплуатации
- Indesit WITL 1067 Инструкция по эксплуатации
- Ada 2D Basic Liner А00239 Инструкция по эксплуатации
- Fluke 566 Инструкция по эксплуатации
- Bosch PHS2102 Инструкция по эксплуатации
- Digitech VOCAL 300 Инструкция по эксплуатации
- LG MS-1724W Инструкция по эксплуатации
- Indesit IWDC 71057 Инструкция по эксплуатации
- Belkin F9A602 Инструкция по эксплуатации
- Geo-Fennel FL 30 271000 Инструкция по эксплуатации