IKA RO 5 0003690500 [13/64] Accessories

IKA RO 5 0003690500 [13/64] Accessories
13
Error Codes (RT 5/10/15)
Accessories
• Stirring bar: ø 8 mm; length 30 mm
RSE Stirring bar remover
Ordering spare parts
When ordering spare parts, please give:
- Device type
- Manufacturing number, see type plate
- Software version
- Item number and designation of the spare part,
see www.ika.com.
Repair
Please send instrument in for repair only after it has been
cleaned and is free from any materials which may consti-
tute a health hazard.
For this you should request the “Safety Declaration (Deconta-
mination Certificate)” from IKA, or use the download printout
of it from the IKA website www.ika.com.
Return the instrument in its original packaging. Storage packaging
is not sucient. Also, please use suitable shipping package mate-
rials.
Error code Cause Effect Solution
Er03 Temperature inside device is too high Heating off - Switch off device and allow to cool down
Er21 Safety relay doesn‘t open during test Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB
Er22 Safety temperature is too low during
test
Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB and safety temperature sensor
Er25 Heating and switching element
monitoring
Heating off - Check the external temperature controller
- Switch off device
- Warning! Only to be carried out by authorised service personnel: Carry
out an internal test on the device to check the plug-in connector for
the heating element, the safety temperature sensor or the PCB
Er26 Difference between temperature of
safety sensor and temperature of
control sensor: control temperature
> (safety temperature + 40 K)
Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel: Carry
out an internal test on the device to check the plug-in connector for
the temperature sensor
Er31 Triac error detection doesn‘t work Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB
Er46 Difference between temperature of
safety sensor and temperature of
control sensor: safety temperature
> (control temperature + 40 K)
Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in connec-
tor for the temperature sensor
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Please contact the service department;
-
Send the device for repair, including a short description of the fault.

Содержание

Скачать