Fluke 715 [4/16] English instruction sheet page 4
715 English Instruction Sheet
Page 4
1. With the Calibrator OFF, press the
O key. P.S.xx is
displayed, where xx is the turn-off time in minutes. OFF
means the power saver is disabled.
2. Press
] and/or } to increase or decrease the turn-
off time in minutes.
3. To disable, press } until the display shows OFF.
HART
TM
Resistor Mode
The calibrator has a user selectable 250 ohm HART
TM
resistor to
facilitate use with HART
TM
communication devices. The resistor
can be switched in/out at any time by simultaneously pressing
the
V and A keys. Use a HART
TM
communicator when
measuring dc mA with loop power or sourcing mA.
Span Check Feature
The calibrator allows the users to store 0% and 100% setpoints
for each output function. Once setpoints are stored, the span
check feature allows the user to quickly toggle back and forth
from 0% to 100% or to step in 25% increments.
Automatic step and ramp modes can be enabled while in the
span check mode by simultaneously pressing the
] } keys.
First select the desired output mode (V,mV or mA) then preceed
to store the setpoints:
1. Storing 0% and 100% Setpoints
2. Using the ] } controls set the output to the desired
output value for 0%.
3. Press the ] } scroll keys simultaneously to store
the 0% value.
4. Now using the
] } controls set the output to the
desired output value for 100%
5. Again, press the ] } scroll keys simultaneously to
store the 100% value.
Содержание
- English instruction sheet page 1 1
- Instruction sheet 1
- Introduction 1
- Volt ma calibrator 1
- English instruction sheet page 2 2
- Auto shut off power saver feature 3
- Caution 3
- English instruction sheet page 3 3
- Safety information 3
- Turning the calibrator on 3
- W warning 3
- English instruction sheet page 4 4
- Resistor mode 4
- Span check feature 4
- English instruction sheet page 5 5
- Measuring dc volts 5
- Sourcing dc volts 5
- English instruction sheet page 6 6
- Measuring dc ma 6
- English instruction sheet page 7 7
- Measuring dc ma with loop power 7
- English instruction sheet page 8 8
- Sourcing ma 8
- Using the current output modes 8
- English instruction sheet page 9 9
- Insert the test leads into the ma simulate and jacks as shown below 9
- Power supply 30v dc max 9
- Simulating a transmitter 9
- Use simulate mode when an external 24 to 30 v loop power supply is available 9
- English instruction sheet page 10 10
- In case of difficulty 10
- Maintenance 10
- Calibration 11
- Cleaning 11
- English instruction sheet page 11 11
- Replacing the battery 11
- Warning 11
- English instruction sheet page 12 12
- Replacement parts and accessories 12
- English instruction sheet page 13 13
- Dc ma input 14
- Dc ma output 14
- Dc v input and output 14
- English instruction sheet page 14 14
- Loop power 14
- Specifications 14
- English instruction sheet page 15 15
- General specifications 15
- English instruction sheet page 16 16
Похожие устройства
- LG MH-6642W Инструкция по эксплуатации
- Belkin F8N309cw Инструкция по эксплуатации
- Geo-Fennel FL 240HV Basic Set 244100 Инструкция по эксплуатации
- Centurion ONE TOUCH Инструкция по эксплуатации
- Bosch PPW2201 Инструкция по эксплуатации
- LG MH-6384BLC Инструкция по эксплуатации
- Fluke 717 Инструкция по эксплуатации
- Belkin F8N335cw011 Инструкция по эксплуатации
- Fluke 718 Инструкция по эксплуатации
- Geo-Fennel FL 40-4Linner 520000 Инструкция по эксплуатации
- LG MH-6384BC Инструкция по эксплуатации
- Centurion XABRE Инструкция по эксплуатации
- Digitech RP100 Инструкция по эксплуатации
- Hotpoint-Ariston ARXF 109 CSI Инструкция по эксплуатации
- Belkin F8N258cwBR Инструкция по эксплуатации
- Fluke 718 ex Инструкция по эксплуатации
- Ada PROF X20 (с поверкой) Инструкция по эксплуатации
- LG MH-6355F Инструкция по эксплуатации
- Centurion XANTA Инструкция по эксплуатации
- Digitech RP200 Инструкция по эксплуатации