Percent Error Setup
Press and hold a. After 3 seconds the set icon and 0% appears on the lower
display. Use
Wand
X
to adjust the 0% point for the Percent Error calculation,
then press D(ENTER) key to confirm selection. Press a. 100% is displayed
on the lower display. Use
Wand
X
to adjust the 100% point for the Percent Error
calculation. Press D (ENTER) to confirm the selection and exit.
Calibrating a P/I Transmitter
To calibrate a P/I (pressure to current) transmitter:
1. Open the pump vent and zero the Calibrator before applying pressure.
Repeat frequently to ensure accuracy.
2. Apply a pressure to the transmitter and measure the transmitter’s current
loop output. OL (overload) is displayed at full scale.
3. Connect the Calibrator to the transmitter as shown in Figure 4 or Figure 5.
The Calibrator recognizes only the pressure module if both types of
connection are in place.
Содержание
- English 1
- Instruction sheet 1
- Pressure calibrators 1
- Series 1
- English 2
- Input units 2
- Symbols 2
- English 3
- Safety 3
- Xw warnings 3
- English 4
- Getting acquainted 4
- Hold in fixed position 4
- Power saver 4
- English 5
- English 6
- Pushbutton functions 6
- Switch test 6
- English 7
- Sourcing loop voltage 7
- Zeroing with absolute pressure modules 7
- Calibrating a p i transmitter 8
- English 8
- Percent error setup 8
- English 9
- Figure 4 connecting to an internal pressure sensor 9
- Min max 9
- Psi 689kpa 9
- S i g n a l 9
- Switch test 9
- T e s t 9
- Calibrating a p i transmitter cont 10
- English 10
- Figure 5 connecting to a pressure module 10
- Ma mode 10
- Min max 10
- Psi 689kpa 10
- S i g n a l 10
- Switch test 10
- T e s t 10
- Calibration 11
- Cleaning 11
- English 11
- In case of difficulty 11
- Maintenance 11
- Replacing the battery 11
- Xw warning 11
- English 13
- Figure 7 replacement parts and accessories 13
- Replacement parts and accessories 13
- 1000g 0 1000 psi 6895 kpa or 6 mpa 2000 psi 13790 kpa or 13 mpa 14
- 100g 12 to 100 psi 83 to 690 kpa 200 psi 1379 kpa or 1 mpa 14
- 1500g 0 1500 psi 10342 kpa or 10 mpa 3000 psi 20684 kpa or 20 mpa 14
- 1g 1 to 1 psi 7 to 7 kpa 5 psi 34 kpa 14
- 300g 12 to 300 psi 83 to 2068 kpa or 2 mpa 375 psi 2586 kpa or 2 mpa 14
- 30g 12 to 30 psi 83 to 207 kpa 60 psi 413 kpa 14
- 500g 0 to 500 psi 3447 kpa or 3 mpa 1000 psi 6895 kpa or 6 mpa 14
- Accuracy is specified for 1 year after calibration at operating temperatures of 10 c to 55 c 14
- English 14
- Figure 7 replacement parts and accessories cont 14
- Model range si range metric max si max metric 14
- Pressure specifications 14
- Specifications 14
- English 15
- General specifications 15
- Contacting fluke 16
- English 16
Похожие устройства
- Belkin F8N335cw011 Инструкция по эксплуатации
- Fluke 718 Инструкция по эксплуатации
- Geo-Fennel FL 40-4Linner 520000 Инструкция по эксплуатации
- LG MH-6384BC Инструкция по эксплуатации
- Centurion XABRE Инструкция по эксплуатации
- Digitech RP100 Инструкция по эксплуатации
- Hotpoint-Ariston ARXF 109 CSI Инструкция по эксплуатации
- Belkin F8N258cwBR Инструкция по эксплуатации
- Fluke 718 ex Инструкция по эксплуатации
- Ada PROF X20 (с поверкой) Инструкция по эксплуатации
- LG MH-6355F Инструкция по эксплуатации
- Centurion XANTA Инструкция по эксплуатации
- Digitech RP200 Инструкция по эксплуатации
- Belkin F8N097eaBR Инструкция по эксплуатации
- Hotpoint-Ariston AQSF 05 S CIS L Инструкция по эксплуатации
- Fluke 719 Инструкция по эксплуатации
- Ada 3D Liner 3V А00132 Инструкция по эксплуатации
- LG MH-6354Y Инструкция по эксплуатации
- Centurion XANADU v.1 Инструкция по эксплуатации
- Digitech RP50PS Инструкция по эксплуатации