Damping Your Measurements
The calibrator normally applies a software filter to
dampen measurements in all functions except
continuity. The specifications assume that damping is
turned on. The damping method is a running average
of the last several measurements. Fluke
recommends that you leave damping on. Turning
damping off may be useful when measurement
response is more important than accuracy or noise
reduction. If you want to turn off damping, press the
More Choices softkey twice, then press the Dampen
softkey so that Off appears. Press Dampen again to
turn damping back on. The default state is
On.
Note
If a measurement falls outside a random noise
window, a new average is started. If damping is
turned off, or until measurements are fully
damped, the
g
symbol is displayed.