8. Rectifying faults
Error messages can occur when
the blood-pressure value is unusually high or low („EE“
appears in the display),
you move or talk during the measurement („EE“ appears
in the display),
the cuff tube is not properly inserted („E1“ appears in the
display),
inflation takes longer than 20 seconds („E1“ appears in
the display),
the inflation pressure is higher than 300 mmHg („E2“
appears in the display),
an error occurs when storing the measured values („E3“
appears in the display),
Exceeding measurement range („Er“ appears in the dis-
play).
In the about cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk. Put the batteries back in if neces-
sary, or else replace them.
9. Specifications
Type: BC 16
Measuring method: non-invasive, oscillometric blood
pressure measurement at the wrist,
for wrist circumferences of
13.5 22 cm
Display range: 0 299 mmHg
Measuring range: systolic: 30 260 mmHg/
diastolic: 30 260 mmHg/
pulse: 40 199 beats/min
Accuracy of the systolic ±3 mmHg/
pressure reading: diastolic: ±3 mmHg/
pulse: ±5 % of the displayed value
Measuring uncertainty: (max. permissible standard devia-
tion according to clinical testing:
systolic: 8 mmHg/
diastolic: 8 mmHg
Power supply: 2 x 1.5 V Micro
(Alkaline Type LR 03)
Classification: Application part Type BF
Ambient conditions: +10 °C – +40 °C max. 40 85 %
atmospheric humidity
Storage/transport: -10 °C – +60 °C max. 10 90 %
atmospheric humidity
Key to symbols: Application part Type BF
Important! Read the operating
instructions
These specification are subject to change without notice for
purpose of improvement.