3. Touch the key 2, the symbol “MANU” will be displayed on the LCD.
4. Touch the key 5 or 7 to select the range you need. There are three modes, including 0-20mA%, 0-24
mA%, 4-20 mA%.
5 Touch the key 10 again to exit.
3.1.18. Sub-display frequency measurement
When the device is under the AC voltage measurement function, the frequency will be displayed in the
sub-display area.
3.1.19. User settings (SET mode)
Power up (load battery) the meter, and then press the key 4 “MAX/MIN” 2 times immediately.
Then we can enter the SET mode:
Set1: Automatic power off time setting.
LCD indicate “SET1” at sub-display.
You can select: OFF(disable this function), 10min, 30min, 60min, 90min, 120min.
Press key 5�or 7� to select the auto power off time.
Finish select, press REL key for save the setting.
Then press Range key to next setting, or power off device.
Set2: The threshold of automatic backlight setting.
LCD indicate “SET2” at sub-display, and “LIGH” at main-display.
You can place device at a dark environment(or cover the light sensor).
Finish select, press REL key for save the setting, LCD show on and backlight on.
Then press Range key to next setting, or power off device.
Set3: The temperature unit setting.
LCD indicate “SET3” at sub-display.
You can select: C (centigrade) or F (Fahrenhite).
Press key 5�or 7� to select the unit.
Finish select, press REL key for save the setting.
Then press Range key to next setting, or power off device.
Set4: The Ambient temperature calibration.
LCD indicate “SET4” at sub-display.
You can calibrate the ambient temperature and CJC temperature.
Press key 5�or 7� to increase or decrease the temperature. Set it same as the ambient temperature.
Short circuit the V-COM terminal.
Finish set, press REL key for save the setting. Wait 10second untill LOCK flash again.
Then power off device.